High-density materials do not always induce artifacts on PET/CT: What is responsible for the difference?
Authors:
- Tomohiro KanetaTKTomohiro Kaneta
- Kentaro TakanamiKTKentaro Takanami
- Yuji WakayamaYWYuji Wakayama
- Akihiro SatoASAkihiro Sato
- Shuichi HiganoSHShuichi Higano
- Hiroshi FukudaHFHiroshi Fukuda
- Shogo YamadaSYShogo Yamada
- Shoki TakahashiSTShoki Takahashi
- 1 June 2007
- journal article
- clinical trial
- Published by Wolters Kluwer Health in Nuclear Medicine Communications
- Vol. 28 (6) , 495-499
Abstract
PET/CT often show increased uptake at sites of high-density materials. However, some materials seldom demonstrate increased uptake on PET/CT, such as the materials used in hip prostheses. We hypothesized that the motion of materials may be crucial for such artifacts. Here, we present representative cases, and validate our hypothesis based on the results of phantom studies. A standard cylinder, 20 cm in diameter, was filled with approximately 37 MBq of 18F-based activity, and a pacemaker was attached to the side of the cylinder. This phantom was placed on the bed with the pacemaker side facing the scanner. PET scans were performed using a Biograph LSO DUO. CT scans were performed first for transmission scans, followed by acquisition of emission scans. The phantom was first scanned (protocol 1). The phantom was then moved about 2 cm closer to the distal edge of the bed just after transmission CT scan, and the emission scan was performed (protocol 2). Homogenous uptake was seen in the cylinder in protocol 1, and there was no visible uptake at the site of the pacemaker. In contrast, a clear hotspot was seen at the site of the pacemaker in protocol 2. The uptake in the cylinder was inhomogeneous; that on the pacemaker side of the cylinder was low, while that on the opposite side was high. High-density materials do not show false increased uptake without motion on PET/CT. Motion of these materials surrounded by radioactive organs may play an important role in inducing false increased uptake on PET/CT.Keywords
This publication has 8 references indexed in Scilit:
- PET/CT Imaging Techniques, Considerations, and ArtifactsJournal of Thoracic Imaging, 2006
- The effect of oral contrast on large bowel activity in FDG-PET/CTAnnals of Nuclear Medicine, 2005
- Impact of metallic dental implants on CT-based attenuation correction in a combined PET/CT scannerEuropean Radiology, 2003
- Artifacts at PET and PET/CT Caused by Metallic Hip Prosthetic MaterialRadiology, 2003
- PET attenuation coefficients from CT images: experimental evaluation of the transformation of CT into PET 511-keV attenuation coefficientsEuropean Journal of Nuclear Medicine and Molecular Imaging, 2002
- Head and neck imaging with PET and PET/CT: artefacts from dental metallic implantsEuropean Journal of Nuclear Medicine and Molecular Imaging, 2002
- 7. Survey of Results of Whole Body Imaging Using the PET/CT at the University of Pittsburgh Medical Center PET FacilityClinical Positron Imaging, 2000
- Attenuation correction for a combined 3D PET/CT scannerMedical Physics, 1998