Can echocardiographic particle image velocimetry correctly detect motion patterns as they occur in blood inside heart chambers? A validation study using moving phantoms
Open Access
- 6 June 2012
- journal article
- research article
- Published by Springer Nature in Cardiovascular Ultrasound
- Vol. 10 (1) , 24
Abstract
Aims: To validate Echo Particle Image Velocimetry (PIV) Methods: High fidelity string and rotating phantoms moving with different speed patterns were imaged with different high-end ultrasound systems at varying insonation angles and frame rates. Images were analyzed for velocity and direction and for complex motion patterns of blood flow with dedicated software. Post-processing was done with MATLAB-based tools (Dflow, JUV, University Leuven). Results: Velocity estimation was accurate up to a velocity of 42 cm/s (r = 0.99, p < 0.001, mean difference 0.4 ± 2 cm/s). Maximally detectable velocity, however, was strongly dependent on frame rate and insonation angle and reached 42 cm/s under optimal conditions. At higher velocities estimates became random. Direction estimates did depend less on velocity and were accurate in 80-90%. In-plane motion patterns were correctly identified with three ultrasound systems. Conclusion: Echo-PIV appears feasible. Velocity estimates are accurate, but the maximal detectable velocity depends strongly on acquisition parameters. Direction estimation works sufficiently, even at higher velocities. Echo-PIV appears to be a promising technical approach to investigate flow patterns by echocardiography.Keywords
This publication has 20 references indexed in Scilit:
- Echocardiographic Particle Image Velocimetry: A Novel Technique for Quantification of Left Ventricular Blood Vorticity PatternJournal of the American Society of Echocardiography, 2010
- Three-dimensional filling flow into a model left ventricleJournal of Fluid Mechanics, 2005
- Flow patterns in the aortic root and the aorta studied with time-resolved, 3-dimensional, phase-contrast magnetic resonance imaging: implications for aortic valve–sparing surgeryThe Journal of Thoracic and Cardiovascular Surgery, 2004
- Accuracy of spectral Doppler flow and tissue velocity measurements in ultrasound systemsUltrasound in Medicine & Biology, 2004
- Birth of three-dimensionality in a pulsed jet through a circular orificeJournal of Fluid Mechanics, 2003
- Three dimensional flow in the human left atriumHeart, 2001
- Left ventricular blood flow patterns in normal subjects: A quantitative analysis by three-dimensional magnetic resonance velocity mappingJACC, 1995
- Particle-Imaging Techniques for Experimental Fluid MechanicsAnnual Review of Fluid Mechanics, 1991
- STATISTICAL METHODS FOR ASSESSING AGREEMENT BETWEEN TWO METHODS OF CLINICAL MEASUREMENTThe Lancet, 1986
- Evaluating doppler devices using a moving string test targetJournal of Clinical Ultrasound, 1982