The Skew Ray Issue in Ocular Aberration Measurement
- 1 June 2006
- journal article
- research article
- Published by Wiley in Optometry and Vision Science
- Vol. 83 (6) , 396-398
Abstract
I investigated the “skew ray” issue with respect to ocular aberration measurement. This arises for aberrometers that can distinguish only radial components of aberrations. Using simulations and Zernike aberrations, I found that rotationally symmetric coefficients are correctly determined but that asymmetric coefficients are underestimated. This underestimation is more pronounced as radial order of the aberrations becomes smaller and as angular dependence of aberrations becomes greater. There is also “leakage” to other orders, and errors are slightly affected by the sampling pattern. Aberrometers that can distinguish only radial components of aberrations do not provide good estimates of asymmetric aberrations.Keywords
Funding Information
- Division of Arctic Sciences (DP0558209)
This publication has 4 references indexed in Scilit:
- Recent advances in measurement of monochromatic aberrations of human eyesClinical and Experimental Optometry, 2005
- Mathematical Model of a Placido Disk Keratometer and Its Implications for Recovery of Corneal TopographyOptometry and Vision Science, 1997
- Axial Curvature and the Skew Ray Error in Corneal TopographyOptometry and Vision Science, 1997
- Corneal Topography Reconstruction Algorithm that Avoids the Skew Ray Ambiguity and the Skew Ray ErrorOptometry and Vision Science, 1997