Comparative cyclic stress-strain and fatigue resistance behavior of electron-beam- and gamma-irradiated ultrahigh molecular weight polyethylene
Open Access
- 23 March 2004
- journal article
- research article
- Published by Wiley in Journal of Biomedical Materials Research Part B: Applied Biomaterials
- Vol. 70B (1) , 152-160
Abstract
Fatigue‐related damage in UHMWPE is one of the main causes of long‐term failure in total joint replacements. Crosslinking ultrahigh molecular weight polyethylene (UHMWPE) by gamma or electron‐beam irradiation, in combination with prior or further thermal treatment, enhances its wear resistance against metallic components in total hip replacements, and eventually in knees. However, little information is available on the fatigue response of this modified UHMWPE. The objective of this study was to compare electron‐beam‐irradiated UHMWPE at 50, 100, and 150 kGy, with the well‐known 25 kGy gamma‐irradiated UHMWPE. Two different cyclic tests were performed under tensile stress, with a zero load ratio, R = 0. First, specimens were subjected to a sinusoidal load cycle at 1 Hz, which provided stress–life curves with the use of a failure criterion based on 12% yield strain. Second, specimens were tested under 50 load cycles at a displacement rate of 15 mm/min, which provided information about the evolution of secant modulus and plastic strain. The incubation period was also analyzed. DSC measurements were carried out to check the crystallization effect of irradiation. According to the results of fatigue resistance there was a crossover behavior between gamma‐ and electron‐beam‐irradiated UHMWPE regarding the applied stress. When the stress was higher than the crossover value, the fatigue resistance of γ‐irradiated samples was higher than electron‐beam‐irradiated ones. When the stress was lower, the fatigue behavior was the opposite. The crossover stress depended on the electron‐beam‐irradiation dose. The clinical relevance of this study lies in an improved knowledge of electron‐beam‐irradiated material under extreme mechanical circumstances, such as fatigue. © 2004 Wiley Periodicals, Inc. J Biomed Mater Res Part B: Appl Biomater 70B: 152–160, 2004Keywords
Funding Information
- Comisión Interministerial de Ciencia y Tecnología (CICYT), Spanish Ministry of Science and Technology (AT 2000-1278)
This publication has 37 references indexed in Scilit:
- Fracture behavior of UHMWPE in non-implanted, shelf-aged knee prostheses after gamma irradiation in airBiomaterials, 2001
- Properties of crosslinked ultra-high-molecular-weight polyethyleneBiomaterials, 2001
- New Polys for Old: Contribution or Caveat?Journal of Bone and Joint Surgery, 2001
- Fatigue of biomaterials: a reviewInternational Journal of Fatigue, 2000
- Enhanced wear performance of highly crosslinked UHMWPE for artificial jointsJournal of Biomedical Materials Research, 2000
- Advances in the processing, sterilization, and crosslinking of ultra-high molecular weight polyethylene for total joint arthroplastyBiomaterials, 1999
- Current Concepts Review - Wear in Total Hip and Knee Replacements*Journal of Bone and Joint Surgery, 1999
- A hip simulator evaluation of different forms of highly cross-linked polyethyleneJournal of Biomechanics, 1998
- Lubrication and wear of ultra-high molecular weight polyethylene in total joint replacementsTribology International, 1998
- The effects of gamma radiation sterilization and ageing on the structure and morphology of medical grade ultra high molecular weight polyethylenePolymer, 1996