Calibration, error analysis, and ongoing measurement process monitoring for mass spectrometry

Authors:
  • SVStephen B. Vardeman
    Stephen B. Vardeman
  • JWJoanne R. Wendelberger
    Joanne R. Wendelberger
  • LWLily Wang
    Lily Wang

  • 1 January 2003
Abstract
We consider problems of quantifying and monitoring accuracy and precision of measurement in mass spectrometry, particularly in contexts where there is unavoidable day-to-day/period-to-period changes in instrument sensitivity. First we consider the issue of estimating instrument sensitivity based on data from a typical calibration study. Simple method-of-moments methods, likelihood-based methods, and Bayes methods based on the one-way random effects model are illustrated. Then we consider subsequently assessing the precision of an estimate of a mole fraction of a gas of interest in an unknown. Finally, we turn to the problem of ongoing measurement process monitoring and illustrate appropriate set-up of Shewhart control charts in this application.
All Related Versions