A new cable-tie-based sternal closure device: infectious considerations†
Authors:
Open Access
- 26 April 2013
- journal article
- research article
- Published by Oxford University Press (OUP) in Interactive CardioVascular and Thoracic Surgery
- Vol. 17 (2) , 219-224
Abstract
To determine the difference in sternal infection and other infectious events between conventional wire and cable-tie-based closure techniques post-sternotomy in a collective of patients after cardiac surgery. The sternal ZipFix™ (ZF) system consists of a biocompatible poly-ether-ether-ketone (PEEK) cable-tie that surrounds the sternum through the intercostal space and provides a large implant-to-bone contact. Between 1 February 2011 and 31 January 2012, 680 cardiac operations were performed via sternotomy at our institution. After the exclusion of operations for active endocarditis and early mortality within 7 days, 95 patients were exclusively closed with ZF and could be compared with 498 who were closed with conventional wires (CWs) during the same period. A multivariable logistic regression analysis, including body mass index, renal impairment and emergency as suspected confounders and inverse propensity weights was performed on the infection rate. Total infection rate was 6.1%, with a total of 36 diagnosed sternal infections (5 in ZF and 31 in CW). Comparing ZF with CW with regard to sternal infection, there is no statistically significant difference related to the device (odds ratio: 0.067, confidence interval: 0.04–9.16, P = 0.72). The propensity modelling provided excellent overlap and the mean propensity was almost the same in both groups. Thus, we have observed no difference in receiving either ZF or CW. No sternal instability was observed with the ZF device, unlike 4/31 patients in the CW group. The overall operation time is reduced by 11 min in the ZF group with identical perfusion and clamping times. Our study underlines a neutral effect of the sternal ZipFix™ system in patients regarding sternal infection. Postoperative complications are similar in both sternal closure methods. The cable-tie-based system is fast, easy to use, reliable and safe.Keywords
This publication has 24 references indexed in Scilit:
- A new cable-tie based sternal closure system: description of the device, technique of implantation and first clinical evaluationJournal of Cardiothoracic Surgery, 2012
- REDIA – Auswirkungen der DRG-Einführung auf die RehabilitationDie Rehabilitation, 2011
- An overview of the objectives of and the approaches to propensity score analysesEuropean Heart Journal, 2011
- Can the Impact of Change of Surgical Teams in Cardiovascular Surgery Be Measured by Operative Mortality or Morbidity? A Propensity Adjusted Cohort ComparisonAnnals of Surgery, 2011
- Mediastinitis After Coronary Artery Bypass Grafting Risk Factors and Long-Term SurvivalThe Annals of Thoracic Surgery, 2010
- Economic aspects of deep sternal wound infections☆European Journal of Cardio-Thoracic Surgery, 2010
- Risk Analysis of Deep Sternal Wound Infections and Their Impact on Long-Term Survival: A Propensity AnalysisAnnals of Plastic Surgery, 2008
- Clinical Outcomes of Noninfectious Sternal Dehiscence After Median SternotomyThe Annals of Thoracic Surgery, 2006
- Potential risk of sternal wiresEuropean Journal of Cardio-Thoracic Surgery, 2004
- Incidence of deep and superficial sternal infection after open heart surgeryA ten years retrospective study from 1981 to 1991European Journal of Cardio-Thoracic Surgery, 1995