Implantable Interface for an Arm Neuroprosthesis

Authors:
  • KIKristin Imenes
    Kristin Imenes
  • LMLuca Marchetti
    Luca Marchetti
  • BHBirgitte Kasin Honsvall
    Birgitte Kasin Honsvall
  • POPer Ohlckers
    Per Ohlckers
  • SRSaad Rabbani
    Saad Rabbani
  • CMCarmen Moldovan
    Carmen Moldovan
  • OIOctavian Ionescu
    Octavian Ionescu
  • EFEduard Franti
    Eduard Franti
  • MDMonica Dascalu
    Monica Dascalu
  • LDLidia Dobrescu
    Lidia Dobrescu
  • DDDragos Dobrescu
    Dragos Dobrescu

Abstract
This paper presents an ongoing development of an arm neuroprosthesis implantable interface consisting of a quadrupole implantable cuff-electrode, a neural analog front-end, a low power Arduino microcontroller, an off the shelf 433MHz transmitter, and a 3.7V 430mAh Li-Po battery. Implantable cuff electrode made of PDMS and Au has been successfully fabricated and tested for biocompatibility. A low power and low noise analog front-end were designed to filter and amplify the electroneurogram signal sensed by the cuff-electrode. The system was tested in the laboratory with input signals similar to the actual biological signals, verifying the correct operation of the implantable device. Finally, the analog front-end module was encapsulated in PDMS and implanted in the hind limb of a swine giving valuable training and knowledge for further development of the neuroprosthesis implantable interface.

This publication has 20 references indexed in Scilit: