Distortion‐corrected image reconstruction with deep learning on an MRI‐Linac
Open Access
- 1 May 2023
- journal article
- research article
- Published by Wiley in Magnetic Resonance in Medicine
- Vol. 90 (3) , 963-977
Abstract
Purpose: MRI is increasingly utilized for image‐guided radiotherapy due to its outstanding soft‐tissue contrast and lack of ionizing radiation. However, geometric distortions caused by gradient nonlinearities (GNLs) limit anatomical accuracy, potentially compromising the quality of tumor treatments. In addition, slow MR acquisition and reconstruction limit the potential for effective image guidance. Here, we demonstrate a deep learning‐based method that rapidly reconstructs distortion‐corrected images from raw k‐space data for MR‐guided radiotherapy applications.Methods: We leverage recent advances in interpretable unrolling networks to develop a Distortion‐Corrected Reconstruction Network (DCReconNet) that applies convolutional neural networks (CNNs) to learn effective regularizations and nonuniform fast Fourier transforms for GNL‐encoding. DCReconNet was trained on a public MR brain dataset from 11 healthy volunteers for fully sampled and accelerated techniques, including parallel imaging (PI) and compressed sensing (CS). The performance of DCReconNet was tested on phantom, brain, pelvis, and lung images acquired on a 1.0T MRI‐Linac. The DCReconNet, CS‐, PI‐and UNet‐based reconstructed image quality was measured by structural similarity (SSIM) and RMS error (RMSE) for numerical comparisons. The computation time and residual distortion for each method were also reported.Results: Imaging results demonstrated that DCReconNet better preserves image structures compared to CS‐ and PI‐based reconstruction methods. DCReconNet resulted in the highest SSIM (0.95 median value) and lowest RMSE (<0.04) on simulated brain images with four times acceleration. DCReconNet is over 10‐times faster than iterative, regularized reconstruction methods.Conclusions: DCReconNet provides fast and geometrically accurate image reconstruction and has the potential for MRI‐guided radiotherapy applications.Keywords
All Related Versions
Funding Information
- National Health and Medical Research Council (1132471)
This publication has 57 references indexed in Scilit:
- Accuracy required and achievable in radiotherapy dosimetry: have modern technology and techniques changed our views?Journal of Physics: Conference Series, 2013
- Parallel MR imagingJournal of Magnetic Resonance Imaging, 2012
- MRI/linac integrationRadiotherapy and Oncology, 2008
- Sparse MRI: The application of compressed sensing for rapid MR imagingMagnetic Resonance in Medicine, 2007
- The fast sinc transform and image reconstruction from nonuniform samples ink-spaceCommunications in Applied Mathematics and Computational Science, 2006
- Investigation of MR image distortion for radiotherapy treatment planning of prostate cancerPhysics in Medicine & Biology, 2006
- A complete distortion correction for MR images: I. Gradient warp correctionPhysics in Medicine & Biology, 2005
- Use of spherical harmonic deconvolution methods to compensate for nonlinear gradient effects on MRI imagesMagnetic Resonance in Medicine, 2004
- Interpolation revisited [medical images application]IEEE Transactions on Medical Imaging, 2000
- Correction of spatial distortion in magnetic resonance angiography for radiosurgical treatment planning of cerebral arteriovenous malformationsMagnetic Resonance Imaging, 1992