Distortion product otoacoustic emissions: Cochlear-source contributions and clinical test performance
- 1 December 2007
- journal article
- conference paper
- Published by Acoustical Society of America (ASA) in The Journal of the Acoustical Society of America
- Vol. 122 (6) , 3539-3553
Abstract
It has been proposed that the clinical accuracy of distortion product otoacoustic emissions (DPOAEs) is affected by the interaction of distortion and reflection sources contributing to the response. This study evaluated changes in dichotomous-decision test performance and threshold-prediction accuracy when DPOAE source contribution was controlled. Data were obtained from 205 normal and impaired ears with ranging from 0 to SPL and and . Data were collected for control conditions (no suppressor, ) and with presented at three levels that previously had been shown to reduce the reflection-source contribution. The results indicated that controlling source contribution with a suppressor did not improve diagnostic accuracy (as reflected by relative operating characteristic curve area) and frequently resulted in poorer test performance compared to control conditions. Likewise, correlations between DPOAE and behavioral thresholds were not strengthened when using the suppressors to control source contribution. While improvements in test accuracy were observed for a subset of subjects (normal ears with the smallest DPOAEs and impaired ears with the largest DPOAEs), the lack of improvement for the larger, unselected subject group suggests that DPOAEs should be recorded in the clinic without attempting to control the source contribution with a suppressor.
Keywords
This publication has 59 references indexed in Scilit:
- Reduced “Audiogram Ripple” in Normally-hearing Subjects with Weak Otoacoustic EmissionsInternational Journal of Audiology, 1999
- Comparison between intensity and pressure as measures of sound level in the ear canalThe Journal of the Acoustical Society of America, 1998
- Suppression of distortion product otoacoustic emissions (DPOAE) near 2f1−f2 removes DP-gram fine structure—Evidence for a secondary generatorThe Journal of the Acoustical Society of America, 1998
- Fine structure of the 2 f1−f2 acoustic distortion product: Effects of primary level and frequency ratiosThe Journal of the Acoustical Society of America, 1997
- Latency and multiple sources of distortion product otoacoustic emissionsThe Journal of the Acoustical Society of America, 1996
- Distortion Product Otoacoustic Emission Test of Sensorineural Hearing Loss: Performance Regarding Sensitivity, Specificity and Receiver Operating CharacteristicsActa Oto-Laryngologica, 1996
- The origin of periodicity in the spectrum of evoked otoacoustic emissionsThe Journal of the Acoustical Society of America, 1995
- Fine structure of the 2f1-f2 acoustic distortion product: Changes with primary levelThe Journal of the Acoustical Society of America, 1993
- Interrelation of different oto-acoustic emissionsThe Journal of the Acoustical Society of America, 1984
- A Ripple Effect in the AudiogramNature, 1958