Dosimetry of cone‐defined stereotactic radiosurgery fields with a commercial synthetic diamond detector
- 30 November 2014
- journal article
- Published by Wiley in Medical Physics
- Vol. 41 (11) , 111702
Abstract
Purpose: Small field x‐ray beam dosimetry is difficult due to lack of lateral electronic equilibrium, source occlusion, high dose gradients, and detector volume averaging. Currently, there is no single definitive detector recommended for small field dosimetry. The objective of this work was to evaluate the performance of a new commercial synthetic diamond detector, namely, the PTW 60019 microDiamond, for the dosimetry of small x‐ray fields as used in stereotactic radiosurgery (SRS).Methods: Small field sizes were defined by BrainLAB circular cones (4–30 mm diameter) on a Novalis Trilogy linear accelerator and using the 6 MV SRS x‐ray beam mode for all measurements. Percentage depth doses (PDDs) were measured and compared to an IBA SFD and a PTW 60012 E diode. Cross profiles were measured and compared to an IBA SFD diode. Field factors, , were calculated by Monte Carlo methods using BEAMnrc and correction factors, , were derived for the PTW 60019 microDiamond detector.Results: For the small fields of 4–30 mm diameter, there were dose differences in the PDDs of up to 1.5% when compared to an IBA SFD and PTW 60012 E diode detector. For the cross profile measurements the penumbra values varied, depending upon the orientation of the detector. The field factors, , were calculated for these field diameters at a depth of 1.4 cm in water and they were within 2.7% of published values for a similar linear accelerator. The corrections factors, , were derived for the PTW 60019 microDiamond detector.Conclusions: The authors conclude that the new PTW 60019 microDiamond detector is generally suitable for relative dosimetry in small 6 MV SRS beams for a Novalis Trilogy linear equipped with circular cones.Keywords
Funding Information
- Australian Research Council (LP110100401)
- Australian Research Council (LP110100401)
This publication has 31 references indexed in Scilit:
- A practical and theoretical definition of very small field size for radiotherapy output factor measurementsMedical Physics, 2014
- Small fields output factors measurements and correction factors determination for several detectors for a CyberKnife® and linear accelerators equipped with microMLC and circular conesMedical Physics, 2013
- A new formalism for reference dosimetry of small and nonstandard fieldsMedical Physics, 2008
- Small fields: Nonequilibrium radiation dosimetryMedical Physics, 2007
- Head scatter factors for small MV photon fields. Part II: The effects of source size and detectorRadiotherapy and Oncology, 2007
- Normobaric hyperoxia—induced improvement in cerebral metabolism and reduction in intracranial pressure in patients with severe head injury: a prospective historical cohort—matched studyJournal of Neurosurgery, 2004
- The value of the PinPoint ion chamber for characterization of small field segments used in intensity-modulated radiotherapyPhysics in Medicine & Biology, 2000
- Assessment of new small-field detectors against standard-field detectors for practical stereotactic beam data acquisitionPhysics in Medicine & Biology, 1999
- Stereotactic radiosurgery for arteriovenous malformations of the brainJournal of Neurosurgery, 1991
- Stereotactic radiosurgery of meningiomasJournal of Neurosurgery, 1991