Accurate and efficient measurement of channelized Hotelling observer‐based low‐contrast detectability on the ACR CT accreditation phantom
Authors:
- 12 November 2022
- journal article
- research article
- Published by Wiley in Medical Physics
- Vol. 50 (2) , 737-749
Abstract
Background: Current CT quality control (QC) for low‐contrast detectability relies on visual inspection and measurement of contrast‐to‐noise ratio (CNR). However, CNR numbers become unreliable when it comes to nonlinear methods, such as iterative reconstruction (IR) and deep‐learning‐based techniques. Image quality metrics using channelized Hotelling observer (CHO) have been validated to be well correlated with human observer performance on phantom‐based and patient‐based tasks, but it has not been widely used in routine CT QC mainly because the CHO calculation typically requires a large number of repeated scans in order to provide accurate and precise estimate of index of detectability (d′).Purpose: The main goal of this work is to optimize channel filters and other CHO parameters and accurately estimate the low‐contrast detectability with minimum number of repeated scans for the widely used American College of Radiology (ACR) CT accreditation phantom so that it can become practically feasible for routine CT QC tests.Methods: To provide a converged d′ value, an ACR phantom was repeatedly scanned 100 times at three dose levels (24, 12, and 6 mGy). Images were reconstructed with two kernels (FBP Br44 and IR Br44‐3). d′ as a function of number of repeated scans was determined for different number of background regions of interest (ROIs), different number of low‐contrast objects, different number of slices per each object, and different channel filter options. A reference d′ was established using the optimized CHO setting, and the bias of d′ was quantified using the d′ calculated from all 100 repeated scans. The variation of d′ at each condition was estimated using a resampling method combining random subsampling among 100 repeated scans and bootstrapping of the ensembles of signal and background ROIs.Results: Optimized parameters in CHO calculation were determined: two background ROIs per object, four objects per low‐contrast object size, nine non‐overlapping slices per object, and a 4‐channel Gabor filter. The bias and uncertainty were estimated at different numbers of repeated scans using these parameters. When only one single scan was used in the CHO calculation, the bias of d′ was below 6.2% and the uncertainty 15.6‐19.6% for the 6, 5, and 4 mm objects, while with three repeated scans the bias was below 2.0% and uncertainty 8.7‐10.9% for the three object sizes.Conclusion: With optimized parameter settings in CHO, efficient and accurate measurement of low‐contrast detectability on the commonly used ACR phantom becomes feasible, which could potentially lead to adoption of CHO‐based low‐contrast evaluation in routine QC tests.Keywords
Funding Information
- National Institutes of Health (U24EB028936, R01EB017095)
This publication has 33 references indexed in Scilit:
- Correlation between model observer and human observer performance in CT imaging when lesion location is uncertainMedical Physics, 2013
- Prediction of human observer performance in a 2‐alternative forced choice low‐contrast detection task using channelized Hotelling observer: Impact of radiation dose and reconstruction algorithmsMedical Physics, 2013
- Addressing Overutilization in Medical ImagingRadiology, 2010
- Image covariance and lesion detectability in direct fan-beam x-ray computed tomographyPhysics in Medicine & Biology, 2008
- ICRU Report 54: Medical imaging—the assessment of image qualityRadiography, 1997
- Lesion detection in structured noiseAcademic Radiology, 1995
- Model observers for assessment of image quality.Proceedings of the National Academy of Sciences, 1993
- Addition of a channel mechanism to the ideal-observer modelJournal of the Optical Society of America A, 1987
- Mathematical description of the responses of simple cortical cells*Journal of the Optical Society of America, 1980
- A four mechanism model for threshold spatial visionVision Research, 1979