Abstract
This study investigates the impact of radiation emitted by irradiated tristructural isotropic (TRISO) nuclear fuel on detector performance during transmission X-ray computed tomography (XCT) imaging. TRISO fuel, a next-generation nuclear fuel form, underwent post-irradiation examination (PIE) using XCT. This work addresses whether high radiation fields from irradiated TRISO fuel, exceeding 2.00 Sv/hr γ on contact, introduce noise or degrade image quality in the CsI-based CMOS detector of the XCT system. Experimental XCT scans were conducted at Idaho National Laboratory using AGR compacts with varying radiation intensities. Frame-averaging techniques and shielding configurations were employed to mitigate potential radiation-induced detector damage and image noise. Results showed minimal signal increase (<0.05% of detector dynamic range) due to radiation, with no significant degradation in image quality. Complementary MCNP simulations confirmed that low energy photons are absorbed by the detector, but almost all high-energy photons pass through the detector without interaction and absorbed dose levels remain well below thresholds known to affect detector performance. These findings demonstrate that XCT can be reliably used for nondestructive examination of highly radioactive TRISO fuel without fear of damaging equipment or significantly impacting image results. PIE of highly radioactive samples enables faster analysis and supports the accelerated deployment of advanced nuclear reactor technologies.