Abstract
Background and Objective The purpose of this investigation was to evaluate: 1. thermal events during laser irradiation @2.1 μm of the pulp; 2. whether these effects are adequately modeled using an empty pulp chamber/root canal. Study Design/Materials and Methods In extracted human teeth, pulpal access was prepared and thermocouples placed 2, 3 mm apical to the center of the irradiation spot. Pulp‐filled or empty pulp chambers were irradiated using a Ho:YAG laser: Spot Size: 1 mm; Power: 1, 2, 3.5, 4.5W; PRR: 5, 12 Hz; Duration: 10 sec. Thermal measurements were repeated 3×. Results Thermal trends did not differ significantly and correlated positively with power (P < 0.01), PRR (P < 0.01), irradiation duration (P < 0.05). No significant difference was determined between temperatures in empty and pulp‐filled chambers at all parameters at 5 Hz and at 1–2W at 12 Hz (P < 0.05, 2‐tailed Student's t‐test). At 12 Hz and >3.5W, pulp chamber temperatures exceeded those in pulpal tissue (P < 0.05). Conclusion Pulp tissues must be present to ensure clinical relevance of thermal measurements. Lasers Surg. Med. 22:46–50, 1998.