Development of a second generation scanning photoemission microscope with a zone plate generated microprobe at the National Synchrotron Light Source
Authors:
- 1 February 1995
- journal article
- research article
- Published by AIP Publishing in Review of Scientific Instruments
- Vol. 66 (2) , 1416-1418
Abstract
We have been developing an instrument that combines the techniques of x‐ray photoelectron spectroscopy and zone plate microfocusing to perform spectromicroscopy. The X1A undulator provides a bright photon source in the soft x‐ray range with a high degree of spatial coherence (a requirement for zone plate focusing). A spherical grating monochromator selects the desired photon energy in the 280–800 eV range. A Fresnel zone plate focuses the beam to a small spot. Photoelectron spectra can be acquired from the small irradiated area with an electron energy analyzer. With the beam focused on the surface and the sample mechanically scanned, element‐specific or chemical‐state‐specific images of the surface can be obtained.Keywords
This publication has 4 references indexed in Scilit:
- Coherent radiation for x-ray imaging—The soft x-ray undulator and the X1A beamline at the NSLSJournal of X-Ray Science and Technology, 1990
- X-ray spectromicroscopy with a zone plate generated microprobeApplied Physics Letters, 1990
- The photoemission spectromicroscope multiple‐application x‐ray imaging undulator microscope (maximum)Journal of Vacuum Science & Technology A, 1990