The Effect of a Global, Subject, and Device-Specific Model on a Noninvasive Glucose Monitoring Multisensor System
Open Access
- 24 April 2015
- journal article
- research article
- Published by SAGE Publications in Journal of Diabetes Science and Technology
- Vol. 9 (4) , 865-872
Abstract
Background: We study here the influence of different patients and the influence of different devices with the same patients on the signals and modeling of data from measurements from a noninvasive Multisensor glucose monitoring system in patients with type 1 diabetes. The Multisensor includes several sensors for biophysical monitoring of skin and underlying tissue integrated on a single substrate. Method: Two Multisensors were worn simultaneously, 1 on the upper left and 1 on the upper right arm by 4 patients during 16 study visits. Glucose was administered orally to induce 2 consecutive hyperglycemic excursions. For the analysis, global (valid for a population of patients), personal (tailored to a specific patient), and device-specific multiple linear regression models were derived. Results: We find that adjustments of the model to the patients improves the performance of the glucose estimation with an MARD of 17.8% for personalized model versus a MARD of 21.1% for the global model. At the same time the effect of the measurement side is negligible. The device can equally well measure on the left or right arm. We also see that devices are equal in the linear modeling. Thus hardware calibration of the sensors is seen to be sufficient to eliminate interdevice differences in the measured signals. Conclusions: We demonstrate that the hardware of the 2 devices worn on the left and right arms are consistent yielding similar measured signals and thus glucose estimation results with a global model. The 2 devices also return similar values of glucose errors. These errors are mainly due to nonstationarities in the measured signals that are not solved by the linear model, thus suggesting for more sophisticated modeling approaches.Keywords
This publication has 36 references indexed in Scilit:
- In Vivo Noninvasive Monitoring of Glucose Concentration in Human Epidermis by Mid-Infrared Pulsed Photoacoustic SpectroscopyAnalytical Chemistry, 2012
- The effect of blood content on the optical and dielectric skin propertiesPhysiological Measurement, 2010
- Cutaneous Blood Perfusion as a Perturbing Factor for Noninvasive Glucose MonitoringDiabetes Technology & Therapeutics, 2010
- Full-field optical coherence tomography for the rapid estimation of epidermal thickness: study of patients with diabetes mellitus type 1Physiological Measurement, 2009
- Multiparameter techniques for non-invasive measurement of blood glucoseSensors and Actuators B: Chemical, 2009
- Non-invasive glucose monitoring in patients with Type 1 diabetes: A Multisensor system combining sensors for dielectric and optical characterisation of skinBiosensors and Bioelectronics, 2009
- Non-invasive glucose monitoring in patients with diabetes: A novel system based on impedance spectroscopyBiosensors and Bioelectronics, 2006
- Noninvasive Glucose SensingAnalytical Chemistry, 2005
- Cell Membrane Response on d-Glucose Studied by Dielectric Spectroscopy. Erythrocyte and Ghost SuspensionsThe Journal of Physical Chemistry B, 2004
- Dielectric spectroscopy study of specific glucose influence on human erythrocyte membranesJournal of Physics D: Applied Physics, 2003