Fast PRF-based MR thermometry using double-echo EPI: in vivo comparison in a clinical hyperthermia setting
- 8 November 2014
- journal article
- research article
- Published by Springer Nature in Magnetic Resonance Materials in Physics, Biology and Medicine
- Vol. 28 (4) , 305-314
Abstract
To develop and test in a clinical setting a double-echo segmented echo planar imaging (DEPI) pulse sequence for proton resonance frequency (PRF)-based temperature monitoring that is faster than conventional PRF thermometry pulse sequences and not affected by thermal changes in tissue conductivity. Four tumor patients underwent between one and nine magnetic resonance (MR)-guided regional hyperthermia treatments. During treatment, the DEPI sequence and a FLASH PRF sequence were run in an interleaved manner to compare the results from both sequences in the same patients and same settings. Temperature maps were calculated based on the phase data of both sequences. Temperature measurements of both techniques were compared using Passing and Bablok regression and the Bland–Altman method. The temperature results from the DEPI and FLASH sequences, on average, do not differ by more than ΔT = 1 °C. DEPI images showed typically more artifacts and approximately a twofold lower signal-to-noise ratio (SNR), but a sufficient temperature precision of 0.5°, which would theoretically allow for a fivefold higher frame rate. The results indicate that DEPI can replace slower temperature measurement techniques for PRF-based temperature monitoring during thermal treatments. The higher acquisition speed can be exploited for hot spot localization during regional hyperthermia as well as for temperature monitoring during fast thermal therapies.Keywords
This publication has 42 references indexed in Scilit:
- Neo-adjuvant chemotherapy alone or with regional hyperthermia for localised high-risk soft-tissue sarcoma: a randomised phase 3 multicentre studyThe Lancet Oncology, 2010
- Noninvasive magnetic resonance thermography of soft tissue sarcomas during regional hyperthermiaCancer, 2006
- Methods and potentials of magnetic resonance imaging for monitoring radiofrequency hyperthermia in a hybrid systemInternational Journal of Hyperthermia, 2005
- Teilkörperhyperthermie mit einem Radiofrequenz-Multiantennen-Applikator unter online Kontrolle in einem 1,5 T MR-TomographenRöFo - Fortschritte auf dem Gebiet der Röntgenstrahlen und der bildgebenden Verfahren, 2004
- Real‐time MR temperature mapping of rabbit liver in vivo during thermal ablationMagnetic Resonance in Medicine, 2003
- A precise and fast temperature mapping using water proton chemical shiftMagnetic Resonance in Medicine, 1995
- Noninvasive MRI Thermometry with the Proton Resonance Frequency (PRF) Method: In Vivo Results in Human MuscleMagnetic Resonance in Medicine, 1995
- STATISTICAL METHODS FOR ASSESSING AGREEMENT BETWEEN TWO METHODS OF CLINICAL MEASUREMENTThe Lancet, 1986
- A New Biometrical Procedure for Testing the Equality of Measurements from Two Different Analytical Methods. Application of linear regression procedures for method comparison studies in Clinical Chemistry, Part IClinical Chemistry and Laboratory Medicine (CCLM), 1983
- RF magnetic field penetration, phase shift and power dissipation in biological tissue: implications for NMR imagingPhysics in Medicine & Biology, 1978