Acute locking in revision total knee arthroplasty due to disengagement of the locking screw
Authors:
- 9 October 2004
- journal article
- case report
- Published by Wiley in Knee Surgery, Sports Traumatology, Arthroscopy
- Vol. 13 (3) , 190-192
Abstract
Acute locking of the joint in a replaced knee joint is very rare. This report describes an acute locking episode of a revised modular total knee replacement, occurring more than 2 years after surgery. A disengaged screw from the modular femoral component had lodged in the joint at the inferior pole of the patella and required urgent arthroscopic removal. There was no subsequent failure of the stem–condylar junction, nor loosening of the femoral component.Keywords
This publication has 5 references indexed in Scilit:
- Failure of the stem–condyle junction of a modular femoral stem in revision total knee arthroplastyThe Journal of Arthroplasty, 2001
- Disengagement of a Locking Screw from a Modular Stem in Revision Total Knee Arthroplasty. A Report of Three Cases*Journal of Bone and Joint Surgery, 1997
- Revision knee arthroplasty. The limits of press fit medullary fixation.1995
- Concerns With Modularity in Total Hip ArthroplastyClinical Orthopaedics and Related Research, 1994
- The role of arthroscopy in the problem total knee replacementArthroscopy, 1990