Characterization of candidate reference materials for bone lead via interlaboratory study and double isotope dilution mass spectrometry
Open Access
- 12 November 2007
- journal article
- research article
- Published by Royal Society of Chemistry (RSC) in Journal of Analytical Atomic Spectrometry
- Vol. 23 (3) , 298-308
Abstract
Four candidate ground bone reference materials (NYS RMs 05-01 through 04), were produced from lead-dosed bovine and caprine sources, and characterized by interlaboratory study. The consensus value (X) and expanded standard uncertainty (UX) were determined from the robust average and standard deviation of the participants’ data for each NYS RM 05-01 through 04. The values were 1.08 ± 0.04, 15.3 ± 0.5, 12.4 ± 0.5, and 29.9 ± 1.1 µg g−1 Pb, respectively. Youden plots of z-scores showed a statistically significant correlation between the results for pairs of NYS RM 05-02 through 04, indicating common sources of between-laboratory variation affecting reproducibility. NYS RM 05-01 exhibited more random variability, affecting repeatability at low concentration. Some participants using electrothermal atomic absorption spectrometry (ETAAS) exhibited a negative bias compared to the all-method consensus value. Other methods used included inductively coupled plasma mass spectrometry (ICP-MS), isotope dilution (ID-) ICP-MS, and ICP atomic (optical) emission spectroscopy (-OES). The NYS RMs 05-01 through 04 were subsequently re-analyzed in house using double ID-ICP-MS to assign certified reference values (C) and expanded uncertainty (UC) of 1.09 ± 0.03, 16.1 ± 0.3, 13.2 ± 0.3 and 31.5 ± 0.7, respectively, indicating a low bias in the interlaboratory data. SRM 1486 Bone Meal was analyzed for measurement quality assessment, obtaining results in agreement with the certified values within the stated uncertainty. Analysis using a primary reference method based on ID-ICP-MS with full quantification of uncertainty calculated according to ISO guidelines provided traceability to SI units.Keywords
This publication has 31 references indexed in Scilit:
- Calibration of laser ablation inductively coupled plasma mass spectrometry for quantitative measurements of lead in boneJournal of Analytical Atomic Spectrometry, 2006
- The occurrence of lead in the bone tissues of Trematomus bernacchii, Terra Nova Bay, Ross SeaAntarctic Science, 2006
- Effect of bone char application on Pb bioavailability in a Pb-contaminated soilEnvironmental Pollution, 2006
- Three decades of in vivo x‐ray fluorescence of lead in boneX-Ray Spectrometry, 2005
- Certification of a fish otolith reference material in support of quality assurance for trace element analysisJournal of Analytical Atomic Spectrometry, 2005
- Effects of acute and chronic oval exposure of lead on blood pressure and bone mineral density in ratsJournal of Inorganic Biochemistry, 2005
- Certification measurement of the cadmium, copper and lead contents in rice using isotope dilution inductively coupled plasma mass spectrometryAnalytica Chimica Acta, 2004
- The Relationship Between Blood Lead, Bone Lead and Child IntelligenceChild Neuropsychology, 2003
- Guidelines for Achieving High Accuracy in Isotope Dilution Mass Spectrometry (IDMS)2002
- Monitoring human exposure to lead: An assessment of current laboratory performance for the determination of blood leadEnvironmental Research, 1992