ClearLLab 10C reagents panel can be applied to analyze paucicellular samples by flow cytometry
Authors:
Open Access
- 18 November 2024
- journal article
- research article
- Published by Wiley in Cytometry Part B: Clinical Cytometry
- Vol. 108 (2) , 128-136
Abstract
The FDA‐approved ClearLLab 10C Reagents Panel (Beckman Coulter) simplified the diagnosis of leukemias and lymphomas by flow cytometry. However, the requirement of using 3 × 106 cells/mL cannot be met for paucicellular samples. Therefore, we tested whether this 10‐color panel can be reliably employed to analyze specimens with low cell concentrations. Serial dilutions of 16 samples (5 normal, 11 abnormal), yielding concentrations ranging from 3.0 × 106 to 0.0469 × 106 cells/mL (64‐fold difference), were stained using the B‐cell and T‐cell panels of the ClearLLab 10C system, and mean fluorescence intensity (MFI) was measured for each antibody. For each cell dilution, the deviation from the value obtained with the FDA‐approved concentration of 3.0 × 106 cells/mL was calculated. The agreement between the highest and lowest cell concentration data was evaluated by the Bland and Altman method, Pearson's and Spearman's correlation analyses, and linear regression. In all patients, the antigen expression pattern was similar at all cell concentrations tested, and the mean deviation of the MFI from the value obtained using 3.0 × 106 cells/mL never exceeded 10% for any of the antibodies. The Bland–Altman method demonstrated the similarity between results obtained with the FDA‐approved cell concentration and a 64‐fold diluted cell suspension, and a high positive correlation was found between MFI acquired under these two conditions. The tests utilizing the lowest density of cells yielded the same patterns of antigen expression in all patients as those performed with the FDA‐approved concentration, documenting a 100% concordance between these two protocols. The ClearLLab 10C panel can reliably determine the expression of markers of leukemias and lymphomas in paucicellular samples containing as little as 0.0469 × 106 cells/mL (64‐fold lower than the FDA‐approved concentration). This finding markedly expands the applicability of the ClearLLab 10C platform in a clinical setting.Keywords
This publication has 24 references indexed in Scilit:
- Prolonged hematologic toxicity following treatment with chimeric antigen receptor T cells in patients with hematologic malignanciesAmerican Journal of Hematology, 2021
- ESCCA/ISCCA protocol for the analysis of cerebrospinal fluid by multiparametric flow‐cytometry in hematological malignanciesCytometry Part B: Clinical Cytometry, 2020
- Diagnostic performance of the ClearLLab 10C B cell tubeCytometry Part B: Clinical Cytometry, 2020
- A multicenter study evaluation of the ClearLLab 10C panelsCytometry Part B: Clinical Cytometry, 2020
- Initial flow cytometric evaluation of the Clearllab lymphoid screenCytometry Part B: Clinical Cytometry, 2017
- Immunophenotyping of Paucicellular SamplesCurrent Protocols in Cytometry, 2014
- Stabilization media increases recovery in paucicellular cerebrospinal fluid specimens submitted for flow cytometry testingCytometry Part B: Clinical Cytometry, 2013
- Titration of Fluorochrome‐Conjugated Antibodies for Labeling Cell Surface Markers on Live CellsCurrent Protocols in Cytometry, 2010
- Quantitative flow cytometry: Inter-laboratory variationCytometry, 1998
- Study of propidium iodide binding to DNA in intact cells by flow cytometryCell Biochemistry and Biophysics, 1990