Neutron Activation Analysis (NAA) is a widely used non-destructive technique for elemental characterization. However, under mixed neutron fields, threshold reactions induced by fast neutrons may generate the same radio-nuclides as those produced through conventional thermal neutron activation by (n,γ) capture, leading to interference phenomena and potentially biasing elemental quantification. The aluminium case is representative of such interferences, notably with silicon, in biological and material science studies. Existing approaches actually determining the interfering contributions require a dedicated additional analysis, either non-destructive (PIXE, Particle Induced X-ray Emission) or chemical-based. In this work, an innovative methodology was applied to quantify interference using the same irradiation as for the NAA. Irradiations were performed using a d-T generator producing fast neutrons at 14.1 MeV. Different neutron fields were obtained using high-density polyethylene (HDPE) moderators, resulting in distinct thermal/fast neutron ratios. Neutron transport simulations were carried out with MCNP6.2. Activities were determined from γ-spectrometry using a HPGe detector. The measured activities were used to determine the masses of pure and mixed Al and Si samples through a system of equations combining measurements performed under different neutron spectra. The resulting masses were in good agreement with the reference values measured prior to irradiation, with uncertainty intervals encompassing the reference masses. The uncertainty analysis found a significant influence of the 28Si(n,p)28Al reaction cross-section, emphasizing the importance of accurate data for interference correction methodologies. The results confirm the capability of the proposed approach to separate activation and interference contributions and demonstrate its potential for improving the reliability of NAA.