Measurement of radiotherapy x‐ray skin dose on a chest wall phantom
Authors:
- K. Y. QuachKQK. Y. Quach
- J. MoralesJMJ. Morales
- M. J. ButsonMBM. J. Butson
- A. B. RosenfeldARA. B. Rosenfeld
- P. E. MetcalfePMP. E. Metcalfe
- 11 July 2000
- journal article
- Published by Wiley in Medical Physics
- Vol. 27 (7) , 1676-1680
Abstract
Sufficient skin dose needs to be delivered by a radiotherapy chest wall treatment regimen to ensure the probability of a near surface tumor recurrence is minimized. To simulate a chest wall treatment a hemicylindrical solid water phantom of 7.5 cm radius was irradiated with 6 MV x‐rays using and fields at 100 cm source surface distance (SSD) to the base of the phantom. A surface dose profile was obtained from 0 to 180°, in 10° increments around the circumference of the phantom. Dosimetry results obtained from radiochromic film (effective depth of 0.17 mm) were used in the investigation, the superficial doses were found to be 28% (of at the 0° beam entry position and 58% at the 90° oblique beam position. Superficial dose results were also obtained using extra thin thermoluminescent dosimeters (TLD) (effective depth 0.14 mm) of 30% at 0°, 57% at 90°, and a metal oxide semiconductor field effect transistor (MOSFET) detector (effective depth 0.5 mm) of 43% at 0°, 62% at 90°. Because the differences in measured superficial doses were significant and beyond those related to experimental error, these differences are assumed to be mostly attributable to the effective depth of measurement of each detector. We numerically simulated a bolus on/bolus off technique and found we could increase the coverage to the skin. Using an alternate “bolus on,” “bolus off” regimen, the skin would receive 36.8 Gy at 0° incidence and 46.4 Gy at 90° incidence for a prescribed midpoint dose of 50 Gy. From this work it is evident that, as the circumference of the phantom is traversed the SSD increases and hence there is an inverse square fluence fall‐off, this is more than offset by the increase in skin dose due to surface curvature to a plateau at about 90°. Beyond this angle it is assumed that beam attenuation through the phantom and inverse square fall‐off is causing the surface dose to reduce.Keywords
This publication has 26 references indexed in Scilit:
- TLD extrapolation for skin dose determination in vivoRadiotherapy and Oncology, 1996
- Radiochromic Solid-State Polymerization Reaction1996
- Effects of package geometry, materials, and die design on energy dependence of pMOS dosimetersIEEE Transactions on Nuclear Science, 1995
- Repair capacity and kinetics of human skin during fractionated radiotherapy: Erythema, desquamation, and telangiectasia after 3 and 5 year's follow-upRadiotherapy and Oncology, 1989
- Investigations on the near surface dose for three 10‐MV x‐ray beam accelerators with emphasis on the reduction of electron contaminationMedical Physics, 1988
- Buildup/surface dose and exit dose measurements for a 6-MV linear acceleratorMedical Physics, 1986
- An investigation into the presence of secondary electrons in megavoltage photon beams (radiotherapy application)Physics in Medicine & Biology, 1983
- The storage and manipulation of radiation dose data in a small digital computerBritish Journal of Radiology, 1974
- Depth dose in skin for obliquely incident60Co radiationBritish Journal of Radiology, 1972
- Electronic Equilibrium and Transition StagesPhysics in Medicine & Biology, 1965