Low-temperature thermometry in high magnetic fields. VI. Industrial-grade Pt resistors above 66 K; Rh–Fe and Au–Mn resistors above 40 K
Authors:
- 1 April 1988
- journal article
- conference paper
- Published by AIP Publishing in Review of Scientific Instruments
- Vol. 59 (4) , 642-645
Abstract
The characteristics of twelve industrial-grade and one precision platinum resistance thermometers for temperatures from 66 to 300 K and in magnetic fields up to 19 T are described. The results from the 13 models from nine suppliers are combined to produce mean values that may be used to correct any of those thermometers for effects of a magnetic field parallel to its long axis. The correctability error at 19 T is less than ±0.3% between 66 and 110 K and less than ±0.1% above 110 K. Magnetoresistance data for Rh–Fe and Au–Mn resistance thermometers are also presented.Keywords
This publication has 6 references indexed in Scilit:
- Stability of industrial-grade platinum resistance thermometers in the range 77 – 273 KCryogenics, 1984
- The use of industrial-grade platinum resistance thermometers between 77 K and 273 KCryogenics, 1983
- Low-temperature thermometry in high magnetic fields. V. Carbon-glass resistorsReview of Scientific Instruments, 1982
- Magnetic-field-induced temperature error of some low-temperature thermometersJournal of Physics E: Scientific Instruments, 1979
- The influence of a magnetic field on an AuMn resistor at low temperaturesCryogenics, 1975
- Low Temperature Thermometry in High Magnetic Fields. II. Germanium and Platinum ResistorsReview of Scientific Instruments, 1971