Influence of Spatial Resolution in Three-dimensional Cine Phase Contrast Magnetic Resonance Imaging on the Accuracy of Hemodynamic Analysis
Open Access
- 1 January 2017
- journal article
- research article
- Published by Japanese Society for Magnetic Resonance in Medicine in Magnetic Resonance in Medical Sciences
- Vol. 16 (4) , 311-316
Abstract
We aim to elucidate the effect of spatial resolution of three-dimensional cine phase contrast magnetic resonance (3D cine PC MR) imaging on the accuracy of the blood flow analysis, and examine the optimal setting for spatial resolution using flow phantoms. The flow phantom has five types of acrylic pipes that represent human blood vessels (inner diameters: 15, 12, 9, 6, and 3 mm). The pipes were fixed with 1% agarose containing 0.025 mol/L gadolinium contrast agent. A blood-mimicking fluid with human blood property values was circulated through the pipes at a steady flow. Magnetic resonance (MR) images (three-directional phase images with speed information and magnitude images for information of shape) were acquired using the 3-Tesla MR system and receiving coil. Temporal changes in spatially-averaged velocity and maximum velocity were calculated using hemodynamic analysis software. We calculated the error rates of the flow velocities based on the volume flow rates measured with a flowmeter and examined measurement accuracy. When the acrylic pipe was the size of the thoracicoabdominal or cervical artery and the ratio of pixel size for the pipe was set at 30% or lower, spatially-averaged velocity measurements were highly accurate. When the pixel size ratio was set at 10% or lower, maximum velocity could be measured with high accuracy. It was difficult to accurately measure maximum velocity of the 3-mm pipe, which was the size of an intracranial major artery, but the error for spatially-averaged velocity was 20% or less. Flow velocity measurement accuracy of 3D cine PC MR imaging for pipes with inner sizes equivalent to vessels in the cervical and thoracicoabdominal arteries is good. The flow velocity accuracy for the pipe with a 3-mm-diameter that is equivalent to major intracranial arteries is poor for maximum velocity, but it is relatively good for spatially-averaged velocity.Keywords
This publication has 22 references indexed in Scilit:
- Magnetic resonance fluid dynamics for intracranial aneurysms—comparison with computed fluid dynamicsActa Neurochirurgica, 2012
- In vivo hemodynamic analysis of intracranial aneurysms obtained by magnetic resonance fluid dynamics (MRFD) based on time-resolved three-dimensional phase-contrast MRINeuroradiology, 2009
- Comparison of hemodynamics of intracranial aneurysms between MR fluid dynamics using 3D cine phase-contrast MRI and MR-based computational fluid dynamicsNeuroradiology, 2009
- Spatial factors for quantifying constant flow velocity in a small tube phantom: comparison of phase-contrast cine-magnetic resonance imaging and the intraluminal Doppler guidewire methodJapanese Journal of Radiology, 2009
- Time‐resolved three‐dimensional phase‐contrast MRIJournal of Magnetic Resonance Imaging, 2003
- The role of the circle of willis in carotid occlusion: Assessment with phase contrast MR angiography and transcranial duplexEuropean Journal of Vascular and Endovascular Surgery, 1995
- Breakthrough of lysozyme through an affinity membrane of cellulose‐cibacron blueAIChE Journal, 1994
- Semiautomated method for noise reduction and background phase error correction in MR phase velocity dataJournal of Magnetic Resonance Imaging, 1993
- MR angiography by multiple thin slab 3D acquisitionMagnetic Resonance in Medicine, 1991
- Intracranial aneurysms: evaluation by MR angiography.American Journal of Roentgenology, 1990