Direct Electrochemistry and Electrocatalytic Behavior of Hemoglobin Entrapped in Ag@C Nanocables/Gold Nanoparticles Nanocomposites Film

Authors:
  • CMChang-Jie Mao
    Chang-Jie Mao
  • JSJi-Ming Song
    Ji-Ming Song
  • HNHe-Lin Niu
    He-Lin Niu
  • SZSheng-Yi Zhang
    Sheng-Yi Zhang
  • RCRong-Jing Cui
    Rong-Jing Cui

Abstract
Direct electrochemistry of hemoglobin (Hb) was successfully fabricated by immobilizing Hb on the nanocomposites containing of [email protected] nanocables and Au nanoparticles (AuNPs) modified glassy carbon electrode (GCE). The immobilized Hb retained its biological activity and shown high catalytic activities to the reduction of H2O2 by circular dicroism (CD) spectrum, fourier transform infrared (FT-IR) spectrum and cyclic voltammetry (CV). Experimental conditions such as scan rate and pH Value were studied and optimized. The results indicated that the resulting biosensor are linear to the concentrations of H2O2 in the ranges of 6.67×10−7−2.40×10−5 M, and the detection limit is 2.02×10−7 M. The electrochemical biosensor has also high stability and good reproducibility.