Clinical Evaluation of a Loop-Mediated Amplification Kit for Diagnosis of Imported Malaria
Authors:
Open Access
- 30 April 2013
- journal article
- research article
- Published by Oxford University Press (OUP) in The Journal of Infectious Diseases
- Vol. 208 (4) , 637-644
Abstract
Background. Diagnosis of malaria relies on parasite detection by microscopy or antigen detection; both fail to detect low-density infections. New tests providing rapid, sensitive diagnosis with minimal need for training would enhance both malaria diagnosis and malaria control activities. We determined the diagnostic accuracy of a new loop-mediated amplification (LAMP) kit in febrile returned travelers. Methods. The kit was evaluated in sequential blood samples from returned travelers sent for pathogen testing to a specialist parasitology laboratory. Microscopy was performed, and then malaria LAMP was performed using Plasmodium genus and Plasmodium falciparum–specific tests in parallel. Nested polymerase chain reaction (PCR) was performed on all samples as the reference standard. Primary outcome measures for diagnostic accuracy were sensitivity and specificity of LAMP results, compared with those of nested PCR. Results. A total of 705 samples were tested in the primary analysis. Sensitivity and specificity were 98.4% and 98.1%, respectively, for the LAMP P. falciparum primers and 97.0% and 99.2%, respectively, for the Plasmodium genus primers. Post hoc repeat PCR analysis of all 15 tests with discrepant results resolved 4 results in favor of LAMP, suggesting that the primary analysis had underestimated diagnostic accuracy. Conclusions. Malaria LAMP had a diagnostic accuracy similar to that of nested PCR, with a greatly reduced time to result, and was superior to expert microscopy.Keywords
This publication has 29 references indexed in Scilit:
- Severe ImportedPlasmodium falciparumMalaria, France, 1996–2003Emerging Infectious Diseases, 2011
- A large proportion of asymptomatic Plasmodium infections with low and sub-microscopic parasite densities in the low transmission setting of Temotu Province, Solomon Islands: challenges for malaria diagnostics in an elimination settingMalaria Journal, 2010
- Mitochondrial DNA Targets Increase Sensitivity of Malaria Detection Using Loop-Mediated Isothermal AmplificationJournal of Clinical Microbiology, 2010
- History of the discovery of the malaria parasites and their vectorsParasites & Vectors, 2010
- Detection of Four Plasmodium Species by Genus- and Species-Specific Loop-Mediated Isothermal Amplification for Clinical DiagnosisJournal of Clinical Microbiology, 2007
- Treatment of Malaria in the United StatesJAMA, 2007
- Towards Complete and Accurate Reporting of Studies of Diagnostic Accuracy: The STARD InitiativeClinical Chemistry, 2003
- Difficulties in the Prevention, Diagnosis, and Treatment of Imported MalariaArchives of internal medicine (1960), 2000
- The rapid manual ParaSight®-F test. A new diagnostic tool for Plasmodium falciparum infectionTransactions of the Royal Society of Tropical Medicine and Hygiene, 1993
- Identification of the four human malaria parasite species in field samples by the polymerase chain reaction and detection of a high prevalence of mixed infectionsMolecular and Biochemical Parasitology, 1993