Improved High-Resolution Ultrasonic Imaging of the Eye
Authors:
- Ronald H. SilvermanRSRonald H. Silverman
- Jeffrey A. KetterlingJKJeffrey A. Ketterling
- Jonathan MamouJMJonathan Mamou
- D. Jackson ColemanDCD. Jackson Coleman
- 1 January 2008
- journal article
- new instrument
- Published by American Medical Association (AMA) in Archives of Ophthalmology (1950)
- Vol. 126 (1) , 94-97
Abstract
Currently, virtually all clinical diagnostic ultrasound systems used in ophthalmology are based on fixed-focus, single-element transducers. High-frequency (≥20-MHz) transducers introduced to ophthalmology during the last decade have led to improved resolution and diagnostic capabilities for assessment of the anterior segment and the retina. However, single-element transducers are restricted to a small depth of field, limiting their capacity to image the eye as a whole. We fabricated a 20-MHz annular array probe prototype consisting of 5 concentric transducer elements and scanned an ex vivo human eye. Synthetically focused images of the bank eye showed improved depth of field and sensitivity, allowing simultaneous display of the anterior and posterior segments and the full lens contour. This capability may be useful in assessment of vitreoretinal pathologies and investigation of the accommodative mechanism.Keywords
This publication has 10 references indexed in Scilit:
- High-Frequency Ultrasonic Imaging of the Anterior Segment Using an Annular Array TransducerOphthalmology, 2007
- Operational verification of a 40-MHz annular array transducerIEEE Transactions on Ultrasonics, Ferroelectrics, and Frequency Control, 2006
- High-frequency ultrasound annular-array imaging. Part I: array design and fabricationIEEE Transactions on Ultrasonics, Ferroelectrics, and Frequency Control, 2006
- Development of a 35-MHz piezo-composite ultrasound array for medical imagingIEEE Transactions on Ultrasonics, Ferroelectrics, and Frequency Control, 2006
- Design and fabrication of a 40-MHz annular array transducerIEEE Transactions on Ultrasonics, Ferroelectrics, and Frequency Control, 2005
- Design and fabrication of annular arrays for high-frequency ultrasoundIEEE Transactions on Ultrasonics, Ferroelectrics, and Frequency Control, 2004
- High-resolution ultrasonic imaging of the posterior segmentOphthalmology, 2004
- A 30-MHz piezo-composite ultrasound array for medical imaging applicationsIEEE Transactions on Ultrasonics, Ferroelectrics, and Frequency Control, 2002
- Clinical Use of Ultrasound BiomicroscopyOphthalmology, 1991
- The Application of Ultrasonic Locating Techniques to OphthalmologyA.M.A. Archives of Ophthalmology, 1958