Evaluation of the Autohaler Actuator: The Effect of Written Patient Instructions on Correct Use
Authors:
- 1 January 1993
- journal article
- research article
- Published by Taylor & Francis in Journal of Asthma
- Vol. 30 (5) , 373-379
Abstract
The Autohaler actuator is a breath-actuated device designed by 3M Pharmaceuticals. The objective of this study was to measure and improve the effectiveness of the device's package insert instructions (PII). Using only the PII for guidance, 5 of 20 (25%) subjects failed to trigger the device. The PII were revised based on the subject's performance and suggestions. Using the revised PII, only 1 of 20 (5%) different subjects failed to trigger the device. These results indicate that relatively minor modifications to instructions can result in significant improvements in patient use. Eighty-five percent of the participants thought the device was easier to use than a metered-dose inhaler.Keywords
This publication has 4 references indexed in Scilit:
- Use and Misuse of Metered-dose Inhalers by Patients with Chronic Lung Disease: A Controlled, Randomized Trial of Two Instruction MethodsAmerican Review of Respiratory Disease, 1989
- Control of Asthma by AerosolsNew England Journal of Medicine, 1986
- Errors in Inhalation Technique and Efficiency in Inhaler Use in Asthmatic ChildrenAllergy, 1986
- Physician Knowledge in the Use of Canister NebulizersChest, 1983