Fabrication of a Glucose Biosensor Based on Inserted Barrel Plating Gold Electrodes
Authors:
- 1 December 2008
- journal article
- research article
- Published by American Chemical Society (ACS) in Analytical Chemistry
- Vol. 81 (1) , 515-518
Abstract
We demonstrate here the application of barrel plating gold electrodes for fabricating a new type of disposable amperometric glucose biosensor. It is prepared by inserting two barrel plating gold electrodes onto an injection molding plastic base followed by immobilizing with a bioreagent layer and membrane on the electrode surface. The primary function of barrel plating is to provide an economical way to electroplate manufactured parts. The manufacture procedure is simple and can increase the fabrication precision for automation in mass production. At the two-electrode system, the detection of glucose is linear up to 800 mg/dL (i.e., 44.5 mM, r2 > 0.99) in pH 7.4 PBS with a sensitivity of 0.71 μA/mM. Excellent sensor-to-sensor reproducibility shows coefficients of variation of only 0.8−1.4% for the detection of 56.5−561.0 mg/dL glucose. In laboratory trials 176 capillary blood samples with a range of 30−572 mg/dL glucose are used to evaluate the clinical application of the biosensor. A good linear correlation is observed between the measured values of the proposed biosensor and laboratory reference. Error grid analysis verifies that the proposed technique is promising in fabricating biosensor strips on a mass scale. As successfully demonstrated by using whole blood glucose as a model analyte, the fabrication technique can extend into other barrel plating noble metal electrodes for various applications.Keywords
This publication has 36 references indexed in Scilit:
- Continuous Blood Glucose Monitoring with a Thin-Film Optical SensorClinical Chemistry, 2007
- Direct glucose determination in blood using a reagentless optical biosensorBiosensors and Bioelectronics, 2007
- Precision and accuracy of two blood glucose meters: FreeStyle Flash versus One Touch UltraAmerican Journal of Health-System Pharmacy, 2006
- An electrochemical investigation of glucose oxidase at a CdS nanoparticles modified electrodeBiosensors and Bioelectronics, 2005
- Accuracy of the EasyTouch blood glucose self-monitoring system: a study of 516 casesClinica Chimica Acta, 2004
- Thick-film electrodes for measurement of superoxide and hydrogen peroxide based on direct protein?electrode contactsAnalytical and Bioanalytical Chemistry, 2004
- Development of a new immunosensor for pesticide detection: a disposable system with liposome-enhancement and amperometric detectionBiosensors and Bioelectronics, 1998
- Design of a Stable Charge Transfer Complex Electrode for a Third-Generation Amperometric Glucose SensorAnalytical Chemistry, 1996
- Fabrication and characterization of a screen-printed, disposable, amperometric cholesterol biosensorThe Analyst, 1994
- Amperometric enzyme-immunosensor based on ferrocene-mediated amplificationBiosensors, 1987