Crosslink Density Is Reduced and Oxidation Is Increased in Retrieved Highly Crosslinked Polyethylene TKA Tibial Inserts
- 1 January 2017
- journal article
- research article
- Published by Wolters Kluwer Health in Clinical Orthopaedics and Related Research
- Vol. 475 (1) , 128-136
Abstract
And oxidative degradation is not fully understood in total knee arthroplasty (TKA). We wished to determine whether crosslink density is reduced at the articular surfaces of retrieved tibial inserts in contact areas exposed to in vivo mechanical stress. Questions/purposes (1) Does polyethylene crosslink density decrease preferentially in regions of the articular surface of thermally stabilized crosslinked polyethylene tibial components exposed to mechanical stress in vivo; and (2) what is the ramification of decreased crosslink density in TKA in terms of accompanying oxidation of the polyethylene? Methods From May 2011 to January 2014, 90 crosslinked polyethylene tibial components were retrieved during revision surgery as a part of a long-standing implant retrieval program. Forty highly crosslinked polyethylene tibial inserts (27 posterior-stabilized designs and 13 cruciate-retaining designs) retrieved for instability (15 cases), stiffness (11), infection (six), aseptic loosening (four), pain (two), and malposition (two) after a mean time of 18 months were inspected microscopically to identify loaded (burnished) and unloaded (unburnished) regions on the articular surfaces. Swell ratio testing was done according to ASTM F2214 to calculate crosslink density and infrared spectroscopy was used according to ASTM F2102 to measure oxidation. Results The region of the tibial insert influenced crosslink density. Loaded surface regions had a mean crosslink density of 0.19 (95% confidence interval [CI], 0.18-0.19) mol/dm3, lower than the other three regions (loaded subsurface, unloaded surface, and unloaded subsurface), which had crosslink densities of 0.21 (95% CI, 0.21-0.22; p < 0.01) mol/dm3. Peak oxidation levels were higher in loaded regions with a mean oxidation index (OI) of 0.67 (95% CI, 0.56-0.78) versus unloaded regions with a mean OI of 0.36 (95% CI, 0.27-0.45; p < 0.01). Peak oxidation levels were higher in annealed samples with a mean OI of 0.66 (95% CI, 0.52-0.81) versus remelted samples with a mean OI of 0.40 (95% CI, 0.34-0.47; p < 0.01). Conclusions The results suggest that the crosslink density decreases and accompanying oxidation is driven predominantly by contact stress conditions. If crosslink density continues to decrease with continued loading over time, crosslinked polyethylene may not provide a clinical advantage over conventional polyethylene in the long term for TKA. Therefore, we will continue to collect longer term retrievals to evaluate mechanical property changes in crosslinked polyethylenes. Clinical Relevance Although we found a decrease in crosslink density and increase in oxidation in the tibial inserts, the degree of oxidation does not suggest, for now, a reason for concern in these early retrievals. The OI values of the tibial inserts in this study were lower than the critical oxidation level (OI > 3) reported in the literature where polyethylene may lose mechanical properties and have the compromised ability to withstand mechanical loading....Keywords
Funding Information
- National Institutes of Health (AR007281)
- Mary and Fred Trump Institute for Implant Analysis
This publication has 33 references indexed in Scilit:
- Highly Cross-Linked Polyethylene May Not Have an Advantage in Total Knee ArthroplastyHSS Journal®: The Musculoskeletal Journal of Hospital for Special Surgery, 2013
- Oxidative properties and surface damage mechanisms of remelted highly crosslinked polyethylenes in total knee arthroplastyInternational Orthopaedics, 2013
- Alternative bearings in total knee arthroplasty: risk of early revision compared to traditional bearingsActa Orthopaedica, 2013
- In vivo oxidation in retrieved highly crosslinked tibial insertsJournal of Biomedical Materials Research Part B: Applied Biomaterials, 2012
- A new mechanism of oxidation in ultrahigh molecular weight polyethylene caused by squalene absorptionJournal of Biomedical Materials Research Part B: Applied Biomaterials, 2011
- History and Systematic Review of Wear and Osteolysis Outcomes for First-generation Highly Crosslinked PolyethyleneClinical Orthopaedics and Related Research, 2011
- Comparison between highly cross-linked and conventional polyethylene in total knee arthroplastyThe Knee, 2009
- In vivo knee loading characteristics during activities of daily living as measured by an instrumented total knee replacementJournal of Orthopaedic Research, 2008
- In Vivo Oxidation of γ-Barrier–Sterilized Ultra–High-Molecular-Weight Polyethylene BearingsThe Journal of Arthroplasty, 2007
- Development of an extremely wear‐resistant ultra high molecular weight polythylene for total hip replacementsJournal of Orthopaedic Research, 1999