Active beam spectroscopy diagnostics for ITER: Present status (invited)
Authors:
- 1 October 2004
- journal article
- conference paper
- Published by AIP Publishing in Review of Scientific Instruments
- Vol. 75 (10) , 3393-3398
Abstract
An overview is given of the present design of the active beam spectroscopy diagnostics for the ITER. Present spatial resolution and signal-to-noise indicate that, in principle, all proposed measurements are possible covering the entire plasma minor radius. Calculations show that the mirror based periscopes and the impurity coating of the first mirror affect the signal strength and the polarization characteristics of the measured spectra having an impact in the measurement accuracy. On-line calibration techniques and methods to access the first mirror status are addressed.Keywords
This publication has 3 references indexed in Scilit:
- Conceptual design of the charge exchange recombination spectroscopy diagnostic for ITERReview of Scientific Instruments, 2003
- Chapter 1: Overview and summaryNuclear Fusion, 1999
- Self-calibrating magnetic field diagnostics in beam emission spectroscopyReview of Scientific Instruments, 1995