Simultaneous analysis of 35 specific antihypertensive adulterants in dietary supplements using LC/MS/MS
Authors:
- 6 November 2016
- journal article
- research article
- Published by Wiley in Biomedical Chromatography
- Vol. 31 (4)
Abstract
A sensitive and specific LC/MS/MS method was developed for the simultaneous analysis of 35 compounds used for treating hypertension as adulterants in dietary supplements. The method was validated for specificity, linearity, accuracy, precision, limit of detection, limit of quantitation, stability and recovery. The limit of detection and limit of quantitation ranged from 0.20 to 20.0 and 0.50 to 60.0 ng/g, respectively. The linearity was good (r2 > 0.999), with intra‐ and interday precision levels of 0.43–7.87% and 0.65–9.95% and the intra‐ and interday accuracies of 84.36–115.82% and 83.78–118.69%, respectively. The stability (relative standard deviation) was <14.75%. The mean recovery was 80.81–117.86% (relative standard deviation <10.00%). Ninety‐seven commercial dietary supplements available in South Korea were analyzed. While none contained detectable amounts of the 35 antihypertensive compounds, the developed LC/MS/MS procedure can be used for routine analysis to monitor illegal adulteration in various forms of dietary supplements.Keywords
This publication has 15 references indexed in Scilit:
- HPLC–MS–MS for the Simultaneous Determination of Atorvastatin and Amlodipine in Plasma of Hypertensive PatientsChromatographia, 2011
- Detection of adulteration of anti-hypertension dietary supplements and traditional Chinese medicines with synthetic drugs using LC/MSFood Additives & Contaminants: Part A, 2010
- Modified Local Straight-Line Screening to Detect Synthetic Drugs in Adulterated Herbal MedicinesApplied Spectroscopy, 2009
- Analysis of six synthetic adulterants in herbal weight-reducing dietary supplements by LC electrospray ionization-MSFood Additives & Contaminants: Part A, 2008
- Development of a solid phase extraction procedure for HPLC–DAD determination of several angiotensin II receptor antagonists in human urine using mixture designTalanta, 2007
- Application of LC–ESI–MS–MS for detection of synthetic adulterants in herbal remediesJournal of Pharmaceutical and Biomedical Analysis, 2006
- Rapid and reliable determination of illegal adulterant in herbal medicines and dietary supplements by LC/MS/MSJournal of Pharmaceutical and Biomedical Analysis, 2006
- Validated HPLC method for determination of amlodipine in human plasma and its application to pharmacokinetic studiesIl Farmaco, 2005
- An HPLC method for the determination of lisinopril in human plasma and urine with fluorescence detectionJournal of Chromatography B, 2004
- Traditional Chinese medicine an approach to scientific proof and clinical validationPharmacology & Therapeutics, 2000