Physical, Performance, and Dosimetric Characteristics of the Δ-Scan 50 Whole-Body/Brain Scanner

Authors:
  • DBDaniel A. Bassano
    Daniel A. Bassano
  • CCCharles C. Chamberlain
    Charles C. Chamberlain
  • JMJames M. Mozley
    James M. Mozley
  • SKStephen A. Kieffer
    Stephen A. Kieffer

Abstract
The .DELTA.-Scan 50 computed tomographic (CT) scanner has the ability to determine X-ray attenuation coefficients with a precision ranging from 0.3-2.6% depending upon such factors as patient size and scan diameter. Resolution is 20-80% greater than the pixel width, with values ranging from 1.2-2.1 mm. The collimating system determines certain characteristics of the image and affects the dose to the patient: doses for typical procedures are on the order of 1 rad, and gonadal doses are in the millirad range. Potential improvements are discussed.