Abstract
Lawrence Livermore National Laboratory is developing magnetic microcalorimeter-based decay energy spectrometry for nuclear materials analysis. Our standard technique uses gold wire bonds to thermally connect large, external absorbers to microcalorimeter sensors, which limits signal rise time and thus the maximum measurable activity. To address these limitations, we have developed a new absorber sensor direct coupling approach. We find that the direct-coupling approach achieves an order of magnitude reduction in signal rise time as compared to our standard method. We have also demonstrated that the direct-coupling technique enables accurate isotopic analysis, with results in agreement with those obtained by mass spectrometry. This report describes standard and direct coupling DES measurement techniques and presents measurement results, including a comparison of signal rise time, signal amplitude, and isotopic analysis between the two configurations.
Funding Information
  • U.S. Department of Energy by Lawrence Livermore National Laboratory (DE-AC52-07NA27344)
  • National Nuclear Security Administration of the Department of Energy
  • Defense Nuclear Nonproliferation Research and Development
  • Office of International Nuclear Safeguards
  • Lawrence Livermore National Laboratory's Laboratory Directed Research
  • Development Program (23-LW-043)

This publication has 12 references indexed in Scilit: