SENSITIVITY OF THE ELECTRONIC SUBTRACTION METHOD IN EVALUATION OF SIMULATED OPTIC DISC CHANGES
Authors:
- 1 April 1982
- journal article
- research article
- Published by Wiley in Acta Ophthalmologica
- Vol. 60 (2) , 293-300
Abstract
Electronic subtraction method can be used in evaluation of glaucomatous optic disc changes from sequential fundus photographs. To determine the resolving power of the subtraction method, minute changes in the diameter of the model optic cup of known dimensions were simulated by changing the magnification, i.e. the refraction, of an artificial test eye. The results show that with an overall magnification of 40 times a two‐dimensional change of 25 μm or more in the optic disc can be detected with the electronic subtraction in optimal conditions. A larger magnification could improve the sensitivity, but at present it is restricted by unlinearity in the imaging system.Keywords
This publication has 5 references indexed in Scilit:
- Photogrammetry experiments with a model eye.British Journal of Ophthalmology, 1980
- DEMONSTRATION OF GLAUCOMATOUS OPTIC DISC CHANGES BY ELECTRONIC SUBTRACTIONActa Ophthalmologica, 1980
- Ein äquidensitometrisches Verfahren zur objektiven Bestimmung der Cup/Disc-RatioOphthalmologica, 1980
- SOME ESSENTIAL OPTICAL FEATURES OF THE ZEISS FUNDUS CAMERAActa Ophthalmologica, 1977
- Simplified Photogrammetry of the Optic DiscArchives of Ophthalmology (1950), 1976