A comparison of metal/metal and ceramic/metal taper‐trunnion modular connections in explanted total hip replacements
Open Access
- 23 June 2021
- journal article
- research article
- Published by Wiley in Journal of Biomedical Materials Research Part B: Applied Biomaterials
- Vol. 110 (1) , 135-143
Abstract
Corrosion and wear are commonly found at the taper‐trunnion connection of modular total hip arthroplasty (THA) explanted devices. While metal/metal (M/M) modular taper‐trunnion connections exhibit more wear/corrosion than ceramic/metal (C/M) modular taper‐trunnion connections, damage is present in both, regardless of material. This study used a combination of assessment techniques including clinical data, visual scoring assessment, optical imaging, profilometry, and x‐ray photoelectron microscopy (XPS), to investigate wear mechanisms and damage features at the modular taper‐trunnion connection of 10 M/M and 8 C/M explanted THAs. No correlation was found between any demographic variable and corrosion wear and assessment scores. All assessment techniques demonstrated that the stem trunnions had more damage than head tapers for both explant groups and agreed that C/M explants had less corrosion and wear compared to M/M explants. However, visual assessment scores differed between assessment techniques when evaluating the tapers and trunnions within the two groups. Profilometry showed an increase (p <.05) in surface roughness for stem trunnions compared to head tapers for both explant groups. X‐ray photoelectron spectroscopy performed on deposits from two M/M explants found chromium and molybdenum carbides beneath the surface while chromium sulfate and aged bone mineral were found on the surface suggesting that the debris is a result of corrosion rather than wear. These results indicate that taper‐trunnion damage is more prevalent for M/M explants, but C/M explants are still susceptible to damage. More comprehensive analysis of damage is necessary to better understand the origins of taper‐trunnion damage.Keywords
Funding Information
- Inova Health System
This publication has 21 references indexed in Scilit:
- A High Prevalence of Corrosion at the Head–Neck Taper with Contemporary Zimmer Non-Cemented Femoral Hip ComponentsThe Journal of Arthroplasty, 2015
- The Relation Between Titanium Taper Corrosion and Cobalt-Chromium Bearing Wear in Large-Head Metal-on-Metal Total Hip ProsthesesJournal of Bone and Joint Surgery, 2014
- The Reliability of a Scoring System for Corrosion and Fretting, and Its Relationship to Material Loss of Tapered, Modular Junctions of Retrieved Hip ImplantsThe Journal of Arthroplasty, 2014
- Massive Pseudotumor in a 28mm Ceramic-Polyethylene Revision THA: A Case ReportReconstructive Review, 2014
- Pseudotumour Formation Due to Tribocorrosion at the Taper Interface of Large Diameter Metal on Polymer Modular Total Hip ReplacementsThe Journal of Arthroplasty, 2013
- Do Ceramic Femoral Heads Reduce Taper Fretting Corrosion in Hip Arthroplasty? A Retrieval StudyClinical Orthopaedics and Related Research, 2013
- Fretting Corrosion and Trunnion Wear—Is it Also a Problem for Sleeved Ceramic Heads?Seminars in Arthroplasty: JSES, 2012
- Metal-on-Metal Local Tissue Reaction Is Associated With Corrosion of the Head Taper JunctionThe Journal of Arthroplasty, 2012
- Characterisation of wear particles produced by metal on metal and ceramic on metal hip prostheses under standard and microseparation simulationJournal of Materials Science: Materials in Medicine, 2006
- Differences in the fretting corrosion of metal–metal and ceramic–metal modular junctions of total hip replacementsJournal of Orthopaedic Research, 2004