Analysis of the Lifetime of Neural Implants Using In Vitro Test Structures
Open Access
Abstract
The aim of this work was to measure the lifetime of neural implant test samples at two different temperatures, using a method that allows the precise measurement of the sample lifetime, further analysis with the use of Weibull statistics, and examination of the applicability of the Van’t Hoff rule. The correct estimation of the lifetime of neural implants is important to avoid preliminary failures, when used in humans. The novelty lies in the precise data due to the measurement approach, the application of the Weibull statistics to neural test samples, and the examination of the Van’t Hoff rule’s applicability to the longevity of polyimide-based neural implant samples. Several samples that consisted of interdigitated gold strands, encapsulated in polyimide were soaked in ringer solution. One batch was soaked at a temperature of 37 °C, and another was soaked at a temperature of 57 °C. Voltage was applied and measured to identify the occurrence of failures. The long-term experiment was stopped after 458 days for the samples at 37 °C and 423 days for the samples at 57 °C, with several samples still being intact at both temperature levels. The time to failure was measured and used to identify the Weibull parameters that would describe the behavior of the samples. The median lifetime of the samples changed from 363 days at 37 °C to 138 days at 57 °C. The scale and shape factor changed from 396 and 3.7 at 37 °C to 138 and 2 at 57 °C, respectively. The measured mean, median times, and Weibull scale factors were lower than expected from the Van’t Hoff rule. The use of the Van’t hoff rule with for accelerated lifetime tests would lead to an estimation of longer lifetimes than realistic. A reaction rate constant around 1.47 appears more appropriate. While a fourfold difference in lifetime would be expected, only a 2.65-fold difference in the median lifetime and a roughly 2.2-fold difference in the mean and Weibull scale factor were observed. The shift of the Weibull shape parameter from 3.7 at 37 °C to 2 at 57 °C with rising temperatures was observed, indicating differences in failure reasons and stronger aging at lower temperatures. The used method is simple to apply and interpret and allows for a precise anticipation of sample lifetimes.
Keywords
Funding Information
- Deutsche Forschungsgemeinschaft (40101397)
This publication has 22 references indexed in Scilit:
- In vitro and in vivo biostability assessment of chronically-implanted Parylene C neural sensorsSensors and Actuators B: Chemical, 2017
- Rapid evaluation of the durability of cortical neural implants using accelerated aging with reactive oxygen speciesJournal of Neural Engineering, 2015
- Long-term reliability of Al2O3and Parylene C bilayer encapsulated Utah electrode array based neural interfaces for chronic implantationJournal of Neural Engineering, 2014
- Non-Hermetic Micropackage for Chronic Implantable SystemsInternational Symposium on Microelectronics, 2013
- Long termin vitrofunctional stability and recording longevity of fully integrated wireless neural interfaces based on the Utah Slant Electrode ArrayJournal of Neural Engineering, 2011
- Long-term in-vivo Investigation of Parylene-C as Encapsulation Material for Neural InterfacesProcedia Engineering, 2011
- In vitro evaluation of the long-term stability of polyimide as a material for neural implantsBiomaterials, 2010
- The insulation performance of reactive parylene films in implantable electronic devicesBiomaterials, 2009
- Accelerated aging for testing polymeric biomaterials and medical devicesMedical Engineering & Physics, 2008
- Rationale for practical medical device accelerated aging programs in AAMI TIR 17Radiation Physics and Chemistry, 2000