A low noise wearable wireless ECG system with body motion cancellation for long term homecare
Authors:
- 1 October 2013
- conference paper
- Published by Institute of Electrical and Electronics Engineers (IEEE)
- Vol. 11, 507-511
Abstract
This paper presents a low noise wearable wireless Electrocardiogram (ECG) system used for long term homecare and noninvasive monitoring three Leads ECG signals. An ac coupling circuit is implemented to remove the dc offset component created by the body movement and employs a high common mode rejection ratio (CMRR) instrumentation amplifier to reduce electromagnetic interference (EMI). The analog front end is connected to the End-Device which samples the three Leads signals with 200 Hz sampling frequency and sends data to a Coordinator on the PC using ZigBee protocol. A QRS and heart rate detection algorithm is implemented on a GUI data acquisition host. The system is evaluated when the body is walking and running. A reasonable performance is achieved even under body motion.Keywords
This publication has 13 references indexed in Scilit:
- A Wearable Cardiac Monitor for Long-Term Data Acquisition and AnalysisIEEE Transactions on Biomedical Engineering, 2013
- The analysis of volatile organic compounds biomarkers for lung cancer in exhaled breath, tissues and cell linesCancer Biomarkers, 2012
- Multi-Model Diagnosis Method for Lung Cancer based on MOS-SAW Breath Detecting e-NoseAIP Conference Proceedings, 2011
- SHIMMER™: An extensible platform for physiological signal capture2010
- An ultra-high input impedance ECG amplifier for long-term monitoring of athletesMedical Devices: Evidence and Research, 2010
- A wearable wireless ECG sensor: a design with a minimal number of parts2010
- Common mode rejection ratio in differential amplifiersIEEE Transactions on Instrumentation and Measurement, 1991
- A Real-Time QRS Detection AlgorithmIEEE Transactions on Biomedical Engineering, 1985
- 60-Hz Interference in ElectrocardiographyIEEE Transactions on Biomedical Engineering, 1973
- A new system of multiple-lead exercise electrocardiographyAmerican Heart Journal, 1966