Reliability of Point-of-Care Hematocrit Measurement During Liver Transplantation
- 1 December 2017
- journal article
- research article
- Published by Wolters Kluwer Health in Anesthesia & Analgesia
- Vol. 125 (6) , 2038-2044
Abstract
BACKGROUND: Although point-of-care (POC) analyzers are commonly used during liver transplantation (LT), the accuracy of hematocrit measurement using a POC analyzer has not been evaluated. In this retrospective observational study, we aimed to evaluate the accuracy of hematocrit measurement using a POC analyzer and identify potential contributors to the measurement error and their influence on mistransfusion during LT. METHODS: We retrospectively collected 6461 pairs of simultaneous intraoperative hematocrit measurements using POC analyzers and laboratory devices during LTs in 901 patients. The agreement of hematocrit measurements was assessed using Bland-Altman analysis for repeated measurements, while the incidence and magnitude of hematocrit measurement error were compared among 16 different laboratory abnormality categories. A generalized estimating equation analysis was performed to identify potential contributors to falsely low-measured POC hematocrit. Additionally, we defined potential “overtransfusion” in the case when POC hematocrit was P = .039), despite similar amount of blood loss. CONCLUSIONS: Hematocrit measured using the POC device tends to be lower than the laboratory hematocrit measured during LT. Commonly encountered laboratory abnormalities during LT include hypoalbuminemia, hypoproteinemia, and hyperglycemia, which may contribute to falsely low-measured POC hematocrit. Careful consideration of these confounders may help reduce overtransfusion that occurs due to falsely low-measured POC hematocrit.Keywords
This publication has 18 references indexed in Scilit:
- Point-of-Care Hemoglobin/Hematocrit Testing: Comparison of Methodology and TechnologyJournal of Cardiothoracic and Vascular Anesthesia, 2016
- Clinical Risk Scoring Models for Prediction of Acute Kidney Injury after Living Donor Liver Transplantation: A Retrospective Observational StudyPLOS ONE, 2015
- Sodium Homeostasis During Liver Transplantation and Correlation with OutcomesAnesthesia & Analgesia, 2014
- Evaluation of point-of-care analyzers’ ability to reduce bias in conductivity-based hematocrit measurement during cardiopulmonary bypassJournal of Clinical Monitoring and Computing, 2013
- The Impact of Intraoperative Transfusion of Platelets and Red Blood Cells on Survival After Liver TransplantationAnesthesia & Analgesia, 2008
- Alloimmunization to Red Blood Cell Antigens Affects Clinical Outcomes in Liver Transplant PatientsLiver Transplantation, 2007
- Evaluation of the blood gas analyzer GEM(r) PREMIER(tm) 3000Clinical Chemistry and Laboratory Medicine (CCLM), 2004
- Evaluation of STAT-CRIT?? Hematocrit Determination in Comparison to Coulter and CentrifugeAnesthesia & Analgesia, 1993
- Colloids and Hematocrit Measurement by ConductivityAnesthesia & Analgesia, 1991
- A Comparison of Conductivity-Based Hematocrit Determinations With Conventional Laboratory Methods in Autologous Blood TransfusionsAnesthesia & Analgesia, 1990