Monitoring of the Critical Meniscus of Very Low Liquid Volumes Using an Optical Fiber Sensor
- 2 June 2020
- journal article
- research article
- Published by Institute of Electrical and Electronics Engineers (IEEE) in IEEE Sensors Journal
- Vol. 20 (20) , 12232-12240
Abstract
A novel sensing system based on single mode optical fiber in reflective configuration has been developed to measure the critical meniscus height (CMH) of low volumes of liquids, which is then used to calculate the contact angle. The sensing system has been designed especially for very low volumes of liquids (e.g. bio-liquids) and the work has demonstrated that measurements are possible with a minimum liquid volume of 5 μL. The sensing system is based on monitoring the spectral variation induced by the difference in the refractive index regions surrounding the fiber tip, at the air-liquid or liquid-liquid interfaces. From the experiments performed in water, (by immersing and extracting the fiber sensor in the liquid sample), it can be concluded that the CMH forming on the fiber decreases as the temperature increases. The change of temperature (in this experiment from 22 to 60 °C) does not influence the CMH of the sample used in the evaluation (P3 mineral oil), giving an indication of its thermal stability. In addition, a fixed fiber was used to measure the variation in the liquid level when another fiber is immersed in the liquid. The error in the liquid level obtained in the work was small, at 0.34 ± 0.04 %. Such a sensor, allowing accurate measurements with very small quantities is especially useful where liquid sample volumes are limited e.g. biologically sourced liquids or specialized, expensive industrial material in the liquid phase.Keywords
This publication has 28 references indexed in Scilit:
- Oligomeric Forms of Insulin Amyloid Aggregation Disrupt Outgrowth and Complexity of Neuron-Like PC12 CellsPLOS ONE, 2012
- Inertial dynamics of air bubbles crossing a horizontal fluid–fluid interfaceJournal of Fluid Mechanics, 2012
- Simultaneous Measurement of Contact Angle and Surface Tension Using Axisymmetric Drop-Shape Analysis-No Apex (ADSA-NA)Langmuir, 2011
- Critical Meniscus Height of Liquids at the Circular Edge of Cylindrical Rods and DisksLangmuir, 2008
- Fiber Optic Sensors2006
- The method of contact angle measurements and estimation of work of adhesion in bioleaching of metalsBiological Procedures Online, 1999
- Influence of Viscosity on Forced and Spontaneous Spreading: Wilhelmy Fiber Studies Including Practical Methods for Rapid Viscosity MeasurementJournal of Colloid and Interface Science, 1998
- How does a droplet spread?Contemporary Physics, 1987
- Static menisci on the outside of cylindersJournal of Fluid Mechanics, 1965
- THE SURFACE TENSION OF BLOOD SERUM, AND THE DETERMINATION OF THE SURFACE TENSION OF BIOLOGICAL FLUIDSJournal of Clinical Investigation, 1929