Accuracy study of a robotic system for MRI‐guided prostate needle placement
- 8 June 2012
- journal article
- research article
- Published by Wiley in The International Journal of Medical Robotics and Computer Assisted Surgery
- Vol. 9 (3) , 305-316
Abstract
Background: Accurate needle placement is the first concern in percutaneous MRI‐guided prostate interventions. In this phantom study, different sources contributing to the overall needle placement error of a MRI‐guided robot for prostate biopsy have been identified, quantified and minimized to the possible extent.Methods: The overall needle placement error of the system was evaluated in a prostate phantom. This error was broken into two parts: the error associated with the robotic system (called ‘before‐insertion error’) and the error associated with needle–tissue interaction (called ‘due‐to‐insertion error’). Before‐insertion error was measured directly in a soft phantom and different sources contributing into this part were identified and quantified. A calibration methodology was developed to minimize the 4‐DOF manipulator's error. The due‐to‐insertion error was indirectly approximated by comparing the overall error and the before‐insertion error. The effect of sterilization on the manipulator's accuracy and repeatability was also studied.Results: The average overall system error in the phantom study was 2.5 mm (STD = 1.1 mm). The average robotic system error in the Super Soft plastic phantom was 1.3 mm (STD = 0.7 mm). Assuming orthogonal error components, the needle–tissue interaction error was found to be approximately 2.13 mm, thus making a larger contribution to the overall error. The average susceptibility artifact shift was 0.2 mm. The manipulator's targeting accuracy was 0.71 mm (STD = 0.21 mm) after robot calibration. The robot's repeatability was 0.13 mm. Sterilization had no noticeable influence on the robot's accuracy and repeatability.Conclusions: The experimental methodology presented in this paper may help researchers to identify, quantify and minimize different sources contributing into the overall needle placement error of an MRI‐guided robotic system for prostate needle placement. In the robotic system analysed here, the overall error of the studied system remained within the acceptable range. Copyright © 2012 John Wiley & Sons, Ltd.Keywords
This publication has 32 references indexed in Scilit:
- A workspace‐orientated needle‐guiding robot for 3T MRI‐guided transperineal prostate intervention: evaluation of in‐bore workspace and MRI compatibilityThe International Journal of Medical Robotics and Computer Assisted Surgery, 2012
- Robotic system for MRI-guided prostate biopsy: feasibility of teleoperated needle insertion and ex vivo phantom studyInternational Journal of Computer Assisted Radiology and Surgery, 2011
- MRI-guided robotic system for transperineal prostate interventions: proof of principlePhysics in Medicine & Biology, 2010
- Integrated navigation and control software system for MRI-guided robotic prostate interventionsComputerized Medical Imaging and Graphics, 2010
- Transperineal prostate biopsy under magnetic resonance image guidance: A needle placement accuracy studyJournal of Magnetic Resonance Imaging, 2007
- Robot-assisted needle placement in open MRI: System architecture, integration and validationComputer Aided Surgery, 2007
- “MRI Stealth” robot for prostate interventionsMinimally Invasive Therapy & Allied Technologies, 2007
- System for prostate brachytherapy and biopsy in a standard 1.5 T MRI scannerMagnetic Resonance in Medicine, 2004
- System for Robotically Assisted Percutaneous Procedures with Computed Tomography GuidanceComputer Aided Surgery, 2001
- System for robotically assisted percutaneous procedures with computed tomography guidanceComputer Aided Surgery, 2001