Rescue Retrieval of AngioGuard Embolic Capture System after Failure of Capture Sheath Retrieval in Carotid Angioplasty and Stenting: Fig 1.
Authors:
Open Access
- 22 May 2008
- journal article
- case report
- Published by American Society of Neuroradiology (ASNR) in American Journal of Neuroradiology
- Vol. 29 (8) , 1594-1595
Abstract
SUMMARY: This report describes the cases of 2 patients with failed retrieval of an AngioGuard distal protection device after carotid stent placement with rescue retrieval by use of a vertebral catheter. An AngioGuard was deployed before carotid angioplasty and stent placement. Attempted removal of the device with the provided recapture sheath was unsuccessful. A 5F 120-cm vertebral catheter was used to recapture and remove the AngioGuard. This technique is a simple and readily available solution for the retrieval of failed removal of a protection device.Keywords
This publication has 3 references indexed in Scilit:
- Protected Carotid-Artery Stenting versus Endarterectomy in High-Risk PatientsNew England Journal of Medicine, 2004
- Endarterectomy for Asymptomatic Carotid Artery StenosisJAMA, 1995
- Beneficial Effect of Carotid Endarterectomy in Symptomatic Patients with High-Grade Carotid StenosisThe New England journal of medicine, 1991