Peripheral dose outside applicators in electron beams
Authors:
- Grigor N GrigorovGGGrigor N Grigorov
- 31 May 2006
- journal article
- Published by IOP Publishing in Physics in Medicine & Biology
- Vol. 51 (12) , N231-N240
Abstract
The peripheral dose outside the applicators in electron beams was studied using a Varian 21 EX linear accelerator. To measure the peripheral dose profiles and point doses for the applicator, a solid water phantom was used with calibrated Kodak TL films. Peak dose spot was observed in the 4 MeV beam outside the applicator. The peripheral dose peak was very small in the 6 MeV beam and was ignorable at higher energies. Using the 10 x 10 cm(2) cutout and applicator, the dose peak for the 4 MeV beam was about 12 cm away from the field central beam axis (CAX) and the peripheral dose profiles did not change with depths measured at 0.2, 0.5 and 1 cm. The peripheral doses and profiles were further measured by varying the angle of obliquity, cutout and applicator size for the 4 MeV beam. The local peak dose was increased with about 3% per degree angle of obliquity, and was about 1% of the prescribed dose (angle of obliquity equals zero) at 1 cm depth in the phantom using the 10 x 10 cm(2) cutout and applicator. The peak dose position was also shifted 7 mm towards the CAX when the angle of obliquity was increased from 0 to 15 degrees.Keywords
This publication has 16 references indexed in Scilit:
- Experimental verification of the application of lateral buildup ratio on the 4-MeV electron beamJournal of Applied Clinical Medical Physics, 2006
- AAPM's TG‐51 protocol for clinical reference dosimetry of high‐energy photon and electron beamsMedical Physics, 1999
- Influence of cone design on the electron beam characteristics on clinical acceleratorsMedical Physics, 1994
- Radiation therapy for skin cancer near the eye: Kilovoltage x-rays versus electronsInternational Journal of Radiation Oncology - Biology - Physics (IJROBP), 1992
- Electron beam modifications for the treatment of superficial malignanciesInternational Journal of Radiation Oncology - Biology - Physics (IJROBP), 1991
- Levels of leakage radiation from electron collimators of a linear acceleratorMedical Physics, 1990
- Radiation leakage through electron applicators on Clinac‐1800 acceleratorsMedical Physics, 1990
- Film dosimetry of clinical electron beamsInternational Journal of Radiation Oncology - Biology - Physics (IJROBP), 1990
- Leakage radiation from electron applicators on a medical acceleratorMedical Physics, 1988
- The in-air scattering of clinical electron beams as produced by accelerators with scanning beams and diaphragm collimatorsPhysics in Medicine & Biology, 1987