Abstract
In an earlier study, we introduced a pelvic ring stability criterion for weightbearing stabilization. In a loading test, however, current external fixation systems alone did not meet this criterion. Internal fixation of the dorsal ring can significantly increase stability, but the condition of severely injured patients is often a contraindication for major surgery. The aim of this study is to optimize external pelvic ring fixation without dorsal ring stabilization to allow weightbearing in early mobilization of patients with unstable pelvic ring injuries. The stiffness of external fixation systems alone and in combination with one or two anterior plates was measured by using a pelvic replica with a type C pelvic ring injury. Endpoints were 15 mm of dislocation or tolerance of 560 N. Addition of one plate at least doubles stiffness, whereas two-plate fixation results in at least a fourfold stiffer configuration. Frame configurations profit more than single-bar systems, and all but one system resist the weightbearing load after double-plating of the pubic symphysis. The choice of double-plate fixation of the anterior ring in addition to external fixation results in weightbearing capacity.Keywords
This publication has 22 references indexed in Scilit:
- Indications and Techniques for External Fixation of the PelvisClinical Orthopaedics and Related Research, 1996
- Vergleichende Untersuchung verschiedener stabilisierungsverfahren bei der Luxation der BeckenhalbgelenkeEuropean Journal of Trauma and Emergency Surgery, 1991
- Upgrading of efficiency in the tracking of body markers with video techniquesMedical & Biological Engineering & Computing, 1991
- Load application to the sacrotuberous ligament; influences on sacroiliac joint mechanicsClinical Biomechanics, 1989
- Comparative Study of the Orthofix and Pittsburgh Frames for External Fixation of Unstable Pelvic Ring FracturesJournal of Orthopaedic Trauma, 1988
- Pelvic ring fractures: should they be fixed?The Journal of Bone and Joint Surgery. British volume, 1988
- Delayed Posterior Internal Fixation of Unstable Pelvic Fractures1987
- Early Open Reduction and Internal Fixation of the Disrupted Pelvic Ring1986
- A Study of External Skeletal Fixation Systems for Unstable Pelvic Fractures1984
- External fixation of the pelvic girdle with a trapezoid compression frameInjury, 1975