Improved Calibration of Instruments for Small Direct Currents
Preprint
Abstract
We report on new calibration methods for picoammeters and low-current sources. The "Ultrastable Low-noise Current Amplifier" (ULCA) was used for the exemplary calibration of commercial state-of-the-art ammeter and current source instruments in the current range between 1 femtoampere and 1 microampere. The uncertainties achieved are compared with results obtained from two other calibration methods for small direct currents. It is shown that using the ULCA as calibrator allows achieving lower uncertainties together with the benefit of easier handling compared to alternative techniques. Also, the ULCA allows performing current meter and source calibrations between 1 femtoampere and 1 microampere with a single method.Keywords
All Related Versions
This publication has 4 references indexed in Scilit:
- Validation of the ultrastable low-noise current amplifier as travelling standard for small direct currentsMetrologia, 2015
- Validation of a quantized-current source with 0.2 ppm uncertaintyApplied Physics Letters, 2015
- Improving the Traceable Measurement and Generation of Small Direct CurrentsIEEE Transactions on Instrumentation and Measurement, 2015
- Ultrastable low-noise current amplifier: A novel device for measuring small electric currents with high accuracyReview of Scientific Instruments, 2015