In vitro assessment of mitral valve function in cyclically pressurized porcine hearts
- 19 February 2016
- journal article
- Published by IOP Publishing in Medical Engineering & Physics
- Vol. 38 (4) , 346-353
Abstract
Recent approaches to the in vitro experimental study of cardiac fluid mechanics involve the use of whole biological structures to investigate in the lab novel therapeutic approaches for the treatment of heart pathologies. To enhance reliability and repeatability, the influence of the actuation strategy of the experimental apparatuses on the biomechanics of biological structures needs to be assessed. Using echography and intracardiac high-speed imaging, we compared the mitral valve (MV) anatomo-functional features (coaptation areas/lengths, papillary muscles-valvular plane distances) in two passive-beating-heart mock loops with internal (IPML) or external (EPML) pressurization of the ventricular chamber. Both apparatuses showed fluid dynamic conditions that closely resembled the physiology. The MVs analyzed in the EPML presented coaptation areas and lengths that were systematically higher, and exhibited greater variability from early-to peak-systole, as compared to those in the IPML. Moreover, in the EPML, the MV leaflets exhibited a convexity with high curvature toward the atrium. With the IPML, MV coaptation lengths ranged similar to available clinical data and the papillary muscles-valve plane distances were more stable throughout systole. In conclusion, both the apparatuses allow for reproducing in vitro the left heart hemodynamics, in terms of flow rates and pressures, with proper mitral valve continence. Results suggest that the IPML is more suitable for replicating the physiological MV functioning, while the EPML may have more potential as a model for the study of MV pathologies.Keywords
This publication has 17 references indexed in Scilit:
- Intracardiac Visualization of Transcatheter Mitral Valve Repair in an In Vitro Passive Beating HeartCirculation, 2015
- Assessment of aortic valve pressure overload and leaflet functions in an ex vivo beating heart loaded with a continuous flow cardiac assist deviceEuropean Journal of Cardio-Thoracic Surgery, 2013
- Intracardiac Visualization of Transcatheter Aortic Valve and Valve-in-Valve Implantation in an In Vitro Passive Beating HeartJACC: Cardiovascular Interventions, 2013
- In vitro hemodynamics and valve imaging in passive beating heartsJournal of Biomechanics, 2012
- In Vitro Characterization of the Mechanisms Responsible for Functional Tricuspid RegurgitationCirculation, 2011
- A Pulsatile Simulator for the in Vitro Analysis of the Mitral Valve with Tri-Axial Papillary Muscle DisplacementThe International Journal of Artificial Organs, 2011
- An Ex Vivo Platform to Simulate Cardiac Physiology: A New Dimension for Therapy Development and AssessmentThe International Journal of Artificial Organs, 2011
- A Novel Approach to the In Vitro Hydrodynamic Study of the Aortic Valve: Mock Loop Development and TestASAIO Journal, 2010
- A Dynamic Heart System to Facilitate the Development of Mitral Valve Repair TechniquesAnnals of Biomedical Engineering, 2009
- Integrated Mechanism for Functional Mitral RegurgitationCirculation, 1997