Evaluation of a New Device for Measuring Responses to Posteroanterior Forces in a Patient Population, Part 1: Reliability Testing
Authors:
Open Access
- 1 February 1996
- journal article
- Published by Oxford University Press (OUP) in Physical Therapy
- Vol. 76 (2) , 158-165
Abstract
Background and Purpose. This study determined the test-retest reliability of measurements of responses to posteroanterior (PA) forces made using a recently developed instrument. Subjects. Twenty-two subjects with nonspecific low back pain participated. Methods. The length of the low-stiffness or “toe“ region and the slope of the linear portion of the force-displacement curve were measured on two occasions. Reliability of repeated measurements was evaluated by intraclass correlation coefficients and percentage-of-agreement scores. The presence of a systematic change in the response to PA forces was evaluated by a paired-samples t test. Results. High reliability coefficients were obtained for both indices of the response to PA forces. The t test revealed no systematic change with repeated testing. Conclusion and Discussion. The new device provides highly reliable measurement of the lumbar PA force-displacement relationship and may prove useful in evaluating theories that have suggested a relationship between spinal symptoms and abnormal PA stiffness.Keywords
This publication has 6 references indexed in Scilit:
- Manipulation: investigation of a proposed mechanismClinical Biomechanics, 1993
- Effect of extensor muscle activation on the response to lumbar posteroanterior forcesClinical Biomechanics, 1993
- Load-displacement-time characteristics of the spine under posteroanterior mobilisationAustralian Journal of Physiotherapy, 1992
- Lumbar Spinal Pathology in Cadaveric Material in Relation to History of Back Pain, Occupation, and Physical LoadingSpine, 1990
- The McGill Pain Questionnaire: Major properties and scoring methodsPain, 1975
- THE ASSESSMENT OF MOBILITY OF METACARPOPHALANGEAL JOINTSRheumatology, 1972