Relationship between Blood Flow in Central Venous Catheters and Hemodialysis Adequacy
Authors:
- Louise M. MoistLMLouise M. Moist
- Brenda R. HemmelgarnBHBrenda R. Hemmelgarn
- Charmaine E. LokCLCharmaine E. Lok
Open Access
- 1 September 2006
- journal article
- research article
- Published by Wolters Kluwer Health in Clinical Journal of the American Society of Nephrology
- Vol. 1 (5) , 965-971
Abstract
Central venous catheter dysfunction is a frequent problem and often is defined as a blood flow <300 ml/min. This prospective, cross-sectional study included 259 patients and examined the relationship between catheter blood flow and dialysis adequacy as measured by urea reduction ratio (URR), single pool urea kinetics, and online effective ionic dialysance clearance. Dialysis adequacy at blood flow rates of <300, <275, and <250 ml/min; sensitivity; specificity; and positive and negative predictive values were calculated. Mean blood flow was 352 ml/min (SD ± 48.8). Mean blood flow <300 ml/min occurred in 10.5% of the patients, and only 26% had a URR of <65%. Maximum blood flows <300 ml/min occurred in 6.9% of patients, and only 22.2% had URR <65%. The positive predictive value of mean blood flow of <300 and <275 ml/min to predict a URR <65% was 22 and 40%, respectively. Using receiver operator characteristic curves, the area under the curve was not significantly different for blood flows of 300, 275, or 250 ml/min. This study indicates that mean blood flows <300 ml/min are not commonly associated with dialysis inadequacy. Setting a single blood flow cut point of <300 ml/min to define the need for intervention will result in a significant number of unnecessary interventions. There is a need to reexamine the definition of catheter dysfunction and expand the definition beyond blood flow rates.Keywords
This publication has 24 references indexed in Scilit:
- Haemodialysis vascular access problems in Canada: results from the Dialysis Outcomes and Practice Patterns Study (DOPPS II)Nephrology Dialysis Transplantation, 2005
- Relationship between urea clearance and ionic dialysance determined using a single-step conductivity profileKidney International, 2005
- Short and long alteplase dwells in dysfunctional hemodialysis cathetersHemodialysis International, 2005
- Efficacy of Tissue Plasminogen Activator for Thrombolysis in Central Venous Dialysis CathetersHemodialysis International, 2003
- Type of vascular access and mortality in U.S. hemodialysis patientsKidney International, 2001
- Relationship between effective ionic dialysance and in vivo urea clearance during hemodialysisAmerican Journal of Kidney Diseases, 2001
- III. NKF-K/DOQI Clinical Practice Guidelines for Vascular Access: Update 2000American Journal of Kidney Diseases, 2001
- Is Ionic Dialysance a Valid Parameter for Quantification of Dialysis Efficiency?Artificial Organs, 1998
- Difference Between Delivered and Prescribed Blood Flow in HemodialysisASAIO Journal, 1996
- Assessing adequacy of hemodialysis: Urea modelingKidney International, 1994