Implementation of free breathing respiratory amplitude‐gated treatments
Open Access
- 13 May 2021
- journal article
- research article
- Published by Wiley in Journal of Applied Clinical Medical Physics
- Vol. 22 (6) , 119-129
Abstract
Purpose: The purpose of this study was to provide guidance in developing and implementing a process for the accurate delivery of free breathing respiratory amplitude‐gated treatments.Methods: A phase‐based 4DCT scan is acquired at time of simulation and motion is evaluated to determine the exhale phases that minimize respiratory motion to an acceptable level. A phase subset average CT is then generated for treatment planning and a tracking structure is contoured to indicate the location of the target or a suitable surrogate over the planning phases. Prior to treatment delivery, a 4DCBCT is acquired and a phase subset average is created to coincide with the planning phases for an initial match to the planning CT. Fluoroscopic imaging is then used to set amplitude gate thresholds corresponding to when the target or surrogate is in the tracking structure. The final imaging prior to treatment is an amplitude‐gated CBCT to verify both the amplitude gate thresholds and patient positioning. An amplitude‐gated treatment is then delivered. This technique was commissioned using an in‐house lung motion phantom and film measurements of a simple two‐field 3D plan.Results: The accuracy of 4DCBCT motion and target position measurements were validated relative to 4DCT imaging. End to end testing showed strong agreement between planned and film measured dose distributions. Robustness to interuser variability and changes in respiratory motion were demonstrated through film measurements.Conclusions: The developed workflow utilizes 4DCBCT, respiratory‐correlated fluoroscopy, and gated CBCT imaging in an efficient and sequential process to ensure the accurate delivery of free breathing respiratory‐gated treatments.Keywords
This publication has 27 references indexed in Scilit:
- The long- and short-term variability of breathing induced tumor motion in lung and liver over the course of a radiotherapy treatmentRadiotherapy and Oncology, 2018
- Planning 4-Dimensional Computed Tomography (4DCT) Cannot Adequately Represent Daily Intrafractional Motion of Abdominal TumorsInternational Journal of Radiation Oncology - Biology - Physics (IJROBP), 2013
- Respiratory gated beam delivery cannot facilitate margin reduction, unless combined with respiratory correlated image guidanceRadiotherapy and Oncology, 2008
- Assessing Respiration-Induced Tumor Motion and Internal Target Volume Using Four-Dimensional Computed Tomography for Radiotherapy of Lung CancerInternational Journal of Radiation Oncology - Biology - Physics (IJROBP), 2007
- The management of respiratory motion in radiation oncology report of AAPM Task Group 76a)Medical Physics, 2006
- Abdominal organ motion measured using 4D CTInternational Journal of Radiation Oncology - Biology - Physics (IJROBP), 2006
- Respiratory correlated cone beam CTMedical Physics, 2005
- Determining parameters for respiration‐gated radiotherapyMedical Physics, 2001
- Deep inspiration breath-hold technique for lung tumors: the potential value of target immobilization and reduced lung density in dose escalationInternational Journal of Radiation Oncology - Biology - Physics (IJROBP), 1999
- Stereotactic Radiotherapy of Malignancies in the Abdomen: Methodological aspectsActa Oncologica, 1994