Evaluating and improving the performance of thin film force sensors within body and device interfaces
- 6 July 2017
- journal article
- research article
- Published by IOP Publishing in Medical Engineering & Physics
- Vol. 48 (1) , 206-211
Abstract
Thin film force sensors are commonly used within biomechanical systems, and at the interface of the human body and medical and non-medical devices. However, limited information is available about their performance in such applications. The aims of this study were to evaluate and determine ways to improve the performance of thin film (FlexiForce) sensors at the body/device interface. Using a custom apparatus designed to load the sensors under simulated body/device conditions, two aspects were explored relating to sensor calibration and application. The findings revealed accuracy errors of 23.3±17.6% for force measurements at the body/device interface with conventional techniques of sensor calibration and application. Applying a thin rigid disc between the sensor and human body and calibrating the sensor using compliant surfaces was found to substantially reduce measurement errors to 2.9±2.0%. The use of alternative calibration and application procedures is recommended to gain acceptable measurement performance from thin film force sensors in body/device applications.Keywords
Funding Information
- Natural Sciences and Engineering Research Council of Canada (493032)
- NSERC (436728-12)
This publication has 19 references indexed in Scilit:
- Force Sensing Resistor and Evaluation of Technology for Wearable Body Pressure SensingJournal of Sensors, 2016
- Evaluation of the force applied by the tongue and lip on the maxillary central incisor toothCoDAS, 2014
- Clinical investigation of the interface pressure in the trans-tibial socket with Dermo and Seal-In X5 liner during walking and their effect on patient satisfactionClinical Biomechanics, 2012
- Measurement of forces applied during Macintosh direct laryngoscopy compared with GlideScope ® videolaryngoscopy*Anaesthesia, 2012
- A Comparison of the Forces Applied to a Manikin during Laryngoscopy with the Glidescope ® and Macintosh LaryngoscopesAnaesthesia and Intensive Care, 2011
- Lip pressure at rest and during function in 2 groups of patients with different occlusionsAmerican Journal of Orthodontics and Dentofacial Orthopedics, 2011
- Evaluation of thin, flexible sensors for time-resolved grip force measurementProceedings of the Institution of Mechanical Engineers, Part C: Journal of Mechanical Engineering Science, 2007
- Forearm Compression by Laparoscopic Hand-Assist DevicesJournal of Endourology, 2004
- Biomechanical Evaluation of the Boston Brace System for the Treatment of Adolescent Idiopathic ScoliosisSpine, 2004
- Evaluation of a sensor for low interface pressure applicationsMedical Engineering & Physics, 2000