Brain tissuepO2-monitoring: Catheterstability and complications

Authors:

Abstract
The authors report on the stability and complications of 73 LFCOX brain ti-p02-microcatheters in 70 patients. Mean monitoring time was 7.5 ± 4.0 days. Patients prone to cerebral hypoxia (after severe head injury (GCS < 9) or a subarachnoid hemorrhage) had a ti-p02-microcatheter inserted next to the fCP-probe in the typical frontal position. After the first 15 insertions, instead of the 3-way-screw (needing a 6 mm burrhole), a l-way-screw (needing a 2.7 mm burrhole) was used for fixation in the bone; by doing so, the procedure can be performed in the.CU and takes only 75 min. Whenever possible a calibration at room air (to determine the sensitivity-drift) and in oxygen free solution (to determine ~he zero-drift) was performed after removal of the catheters. fdeally the expected p02 at room air was around 754 mmHg (temperature dependent) and at zero calibration 0 mmHg. Mean sensitivity-drift for 54 catheters was -8.5 ± 15.4%. Dividing the catheters into groups, depending on the duration of monitoring (7 -4, 5-8 and 9-76 days)/· revealed that the greatest part of the (negative) sensitivity-drift occurred during day 1-4 after insertion. After 1 week of monitoring sometimes a positive drift occurred (being far less than the negative drift durirjg the· first 4 days). Compared to the old catheters (-10.3 ± 17.3%) (on the first half-of the patients) the ·n·w ones. showed a lower sensitivity-drift (- 6.8 ± 13.4%). The zero-drift of 56 catheters was low with mean drift after ·7.5 ± 4.0 days of 1.5 ± 7.5 mmHg. Here also the highest drift occurred on day 7-4 after Jnsertion. No infection was seen and 2 times (2.7%) a small hematomal not needing evacuation occurred. As the ti-p02-catheter (having a smaller diameter) and the fCP-catheter. were inserted at the same time, one cannot distinguish which catheter caused the hematoma. A possible explanation for the occurrence of the two hematomas is the insertion of the catheters too close to the midline. The authors conclude that LFCOX ti-p02-monitoring is a safe and reliable method. Further decrease of the complication rate and increase of the catheter-stability may be expected. [Neural Res 1997; 19: 241-245]